Introduction

Patients with a history of anterior cruciate ligament reconstruction (ACLR) present unique challenges in total knee arthroplasty (TKA) due to altered knee kinematics, potential residual instability, and the presence of fixation implants that may interfere with implant positioning [1, 2]. Addressing these complexities requires a precise and individualized approach. Image-based robotic-assisted TKA had proven to be a safe, reliable, and accurate technique that enables the functional alignment (FA) strategy that accounts for each patient’s bony anatomy, ligamentous laxities, and anterior compartment characteristics to optimize implant placement and knee balance [37]. By leveraging robotic technology, surgeons can achieve a three-dimensional alignment that preserves native knee kinematics while accommodating prior ACLR-related changes.

There is limited guidance on managing the unique challenges of TKA in patients with prior ACLR, particularly regarding hardware interference and altered knee kinematics. This technique illustrates how image-based robotic-assisted TKA with FA can address these issues. A cohort of patients is also presented to support the feasibility of this approach in clinical practice.

Methods

A consecutive cohort of patients with previous ACLR undergoing image-based TKA with the FA principles were evaluated retrospectively from June 2021 to December 2022. Demographics are presented in Table 1, while the Knee Society Score (KSS)-knee and -function were assessed pre- and post-operatively to validate the presented technique.

Table 1

The demographics of the studied cohort are presented. Data are reported as mean values with standard deviations (SD).

The study was approved by the Scientific Committee of Hygeia Hospital (Ref. No. 663, 20/12/2023) and conducted in accordance with the Declaration of Helsinki.

Surgical technique

The patient is placed in a standard supine position, with one arm resting on a lateral support in abduction and the other on the surgical table. The leg is positioned in a leg holder that allows unrestricted movement from full extension to full flexion (Video 1).

FA workflow

The FA technique aims to achieve a balanced three-dimensional implant positioning by considering the patient’s unique bony anatomy, ligamentous laxities, and anterior compartment characteristics [8, 9].

After joint exposure, checkpoints are placed, and anatomical reference points are registered. Real-time intraoperative assessment of coronal alignment and soft tissue balance is performed at full extension, 90° flexion, and maximum flexion. Ligamentous laxities are evaluated using varus and valgus stress maneuvers in extension and in flexion. Adjustments to femoral and tibial cuts are made to achieve optimal balance. The anterior compartment is assessed to maintain native trochlear morphology and avoid overstuffing.

When utilizing an image-based robotic platform, the FA workflow begins with CT-based preoperative planning, where femoral and tibial implant models are digitally superimposed onto the patient’s CT images.

Coronal alignment is planned to maintain native epiphyseal orientation by equalizing medial and lateral resections and correcting for bone wear as needed. We follow a “restricted” FA strategy with the tibial component’s coronal positioning being set within a safe range of 3° varus to 2° valgus (with reference to the femoral mechanical axis), depending on the patient’s overall limb alignment. Sagittal positioning is determined by matching the individual’s native posterior tibial slope, with adjustments based on the selected implant type. For cruciate-retaining (CR) designs, the posterior slope can be set between 0° and 3°, depending on patient-specific anatomy, while for posterior-stabilized (PS) designs, it is typically maintained between 0° and 1° to optimize kinematics and implant stability. Axially, the tibial component is aligned with the tibial anteroposterior axis, ensuring optimal medio-lateral and AP bone coverage.

For the femoral component, coronal alignment is planned within a range of 3° valgus to 3° varus (with reference to the tibial mechanical axis), preserving the patient-specific epiphyseal orientation. Distal and posterior femoral resections are set to maintain the medial joint line height in both flexion and extension while accounting for implant thickness. In the axial plane, femoral rotation is optimized by referencing the transepicondylar axis (TEA) or the posterior condylar axis (PCA), with strict attention to reproducing the patient’s native trochlear groove as visualized on the preoperative CT scan. While no strict rotational boundaries are enforced, accurate trochlear alignment is crucial to avoid mismatch. Sagittal positioning is refined to prevent anterior notching, which can be mitigated by increasing femoral flexion or anterior translation when necessary. The limit for combined flexion (tibial slope and femoral flexion) is 10°.

Intraoperatively, joint balancing is performed using a digital tensioning device to assess soft tissue laxity in both extension and flexion under controlled stress. Adjustments to femoral and tibial component alignment are made in all three dimensions to achieve optimal balance within predefined parameters. The goal is to achieve flexion and extension gaps within 0–1.5 mm, depending on whether a CR or PS design is used, while maintaining an additional 1–2 mm of lateral compartment laxity in flexion to replicate natural knee mechanics. Final implant selection (CR or PS) is determined based on posterior cruciate ligament integrity and intraoperative flexion stability, with trial components used to confirm alignment before definitive implantation.

Image-based guidance and the detection of fixation implants

Image-based robotic-assisted TKA provides a precise method for planning and executing bone resections while accommodating pre-existing hardware from ACLR. The preoperative CT scan allows for the identification of fixation implants, such as interference screws or staples, and aids in strategizing component positioning to minimize disruption of existing structures. Intraoperatively, real-time navigation ensures accurate resections and adjustments, reducing the risk of excessive bone loss or implant malpositioning.

In this specific case, the tibial fixation staple was completely covered by bone, making it difficult to locate through direct visualization. Using the robotic system’s probe, it was identified position intraoperatively (Figure 1). To avoid removing the staple and creating an osseous defect, the tibial and femoral components were moved proximally by 1–2 mm, thereby preserving bone integrity while maintaining optimal alignment and stability. This strategic modification ensured that the implant positioning remained within safe biomechanical limits while avoiding unnecessary hardware removal. The rotation of the implant is verified with the probe with reference to Akagi’s line (Figure 2).

thumbnail Figure 1

Intraoperative identification of the tibial fixation staple using the image-based robotic system. The robotic probe is utilized to locate the fixation staple, which was entirely covered by bone and not visible through direct visualization. The system provides real-time spatial feedback, displaying the distance to the bone and mapping reference points for accurate positioning. The inset image on the right shows the corresponding CT scan with the staple’s location, ensuring precise intraoperative identification.

thumbnail Figure 2

Intraoperative verification of tibial component rotation using robotic guidance. The rotation of the tibial implant is assessed with reference to Akagi’s line. The use of the probe ensures accurate rotational positioning before final implantation.

For the femoral component, preoperative planning with the robotic system allowed to precisely position the cutting box at the head of the femoral fixation screw (Figure 3). Before proceeding with the box cut, the probe was used intraoperatively to confirm the mediolateral positioning of the femoral component, ensuring that the screw head would be accurately located during the cut. This step was crucial in preventing blind bone removal and allowed for a controlled, precise resection that avoided unnecessary bone loss (Figure 4).

thumbnail Figure 3

Preoperative planning of the femoral component and boxing cut using an image-based robotic system. CT-based imaging was utilized to precisely position the femoral implant while ensuring the cutting box aligned with the head of the femoral fixation screw. The robotic system provided detailed visualizations in transverse, coronal, and sagittal planes.

thumbnail Figure 4

Intraoperative verification of the mediolateral positioning of the femoral implant using a robotic-assisted system. The robotic probe was utilized to confirm the precise mediolateral placement of the femoral component before proceeding with the box cut. This step ensured accurate localization of the femoral fixation screw head, preventing blind bone removal and allowing for a controlled resection while preserving bone stock.

Dynamic evaluation of the anterior compartment

Dynamic evaluation of patellar tracking can be performed using the probe, allowing for real-time adjustments to optimize mediolateral tracking and anterior offset as needed [5, 10].

Results

Table 2 presents the outcomes of the studied cohort. There was significant improvement of the KSS-knee and -function at final follow-up. No revisions or complications were recorded in this sample.

Table 2

The preoperative and postoperative evaluation with the Knee Society Score (KSS) is presented. Data are reported as mean values with standard deviations (SD). Independent t-test was used for the comparison of the mean values. Statistical significance was defined as P < 0.05. In italics the statistically significant values.

Discussion

The aim of this surgical technique was to describe the use of image-based robotic-assisted TKA in patients with a history of ACLR, integrating FA principles to optimize implant positioning and knee balance. Furthermore, a cohort of 25 patients undergoing this technique with favorable outcomes at a mean follow-up of 35.44 months was presented.

Unlike systematic alignment approaches, FA focuses on intraoperative assessment of patient-specific ligamentous laxities, allowing for individualized implant positioning that accommodates the native soft tissue envelope [8, 9, 11]. This technique provides greater adaptability in balancing flexion and extension gaps without relying on predefined coronal or sagittal alignment targets, which may lead to suboptimal outcomes in post-ACLR knees. Furthermore, the use of image-based guidance allows for precise localization of pre-existing fixation implants, enabling surgeons to either target selective extraction of interfering hardware or adjust implant positioning to avoid unnecessary removal. This flexibility helps preserve bone stock, maintain implant stability, and reduce the risk of intraoperative complications related to implant fixation interference. The possible advantages and disadvantages of the technique are presented in Table 3.

Table 3

Advantages and disadvantages of the use of the image-based robotic system in total knee arthroplasty in patients with previous anterior cruciate ligament reconstruction.

FA is a relatively new technique but there are short-term studies that have demonstrated that FA in robotic-assisted TKA achieves safe coronal alignment, improved soft tissue balance, and favorable clinical outcomes [1214]. FA has been shown to reduce the need for ligament releases while maintaining joint stability, even in complex knee morphotypes. Given that patients with a history of ACLR often experience altered knee kinematics, residual instability, and patellofemoral dysfunction, applying FA in this population could lead to more physiological post-operative outcomes compared to traditional alignment methods [1, 2]. Furthermore, FA has demonstrated comparable functional and clinical outcomes between different polyethylene insert types, highlighting its adaptability across various knee conditions [15]. FA’s soft tissue-driven approach may optimize function in post-ACLR knees by balancing stability and constraint, addressing their unique biomechanical challenges. Further studies are needed to validate these findings and assess long-term outcomes in this specific patient population.

Conclusion

Image-based robotic-assisted TKA with FA offers a tailored approach for post-ACLR patients by addressing ligamentous laxities and hardware interference, potentially improving outcomes through optimized balance and preservation of native knee mechanics.

Funding

This research received no external funding.

Conflicts of interest

Authors 1, 2, 3, and 4 have nothing to declare. Author 5: Consultant for Stryker. Author 6: Royalties from Smith Nephew, Stryker and Serf; Consultant for Stryker, Heraeus; Institutional research support from Amplitude and Groupe Lepine; Editorial Board for Journal of Bone and Joint Surgery (Am). Author 7: Consultant for Stryker.

Data availability statement

Data is available upon reasonable request to the corresponding author

Author contribution statement

Author 1: Conceptualization, Methodology, Data curation, Writing an original draft.

Author 2: Conceptualization, Methodology, Data curation, Writing an original draft.

Author 3: Data curation, Methodology, Writing, Reviewing, and Editing.

Author 4: Data curation, Methodology, Writing

Author 5: Conceptualization, Methodology, Writing, Reviewing.

Author 6: Conceptualization, Supervision, Validation, Writing, Reviewing, and Editing.

Author 7: Conceptualization, Supervision, Validation, Writing, Reviewing, and Editing.

Ethics approval

Ethical approval was obtained.

Informed consent

Written informed consent was obtained from the patient depicted in the technique presentation.

Supplementary material

Video 1. Step-by-step surgical technique demonstrating the use of an image-based robotic system combined with a functional alignment strategy in patients with a history of anterior cruciate ligament (ACL) reconstruction. Access here

References

  1. Alessio-Mazzola M, Biavardi N, Solarino G, et al. (2024) Total knee arthroplasty after anterior cruciate ligament reconstruction: a narrative review. Ann Jt 9, 25. [CrossRef] [PubMed] [Google Scholar]
  2. Wilson JM, Markos JR, Krych AJ, et al. (2023) Total knee arthroplasty in patients who had a prior anterior cruciate ligament reconstruction: balancing remains the issue. J Arthroplasty 38, S71–S76. [CrossRef] [PubMed] [Google Scholar]
  3. Dretakis K, Koutserimpas C (2024) Pitfalls with the MAKO robotic-arm-assisted total knee arthroplasty. Medicina (Kaunas) 60, 262. [CrossRef] [PubMed] [Google Scholar]
  4. Diquattro E, Prill E, Salzmann M, et al. (2024) High three-dimensional accuracy of component placement and lower limb alignment using a robotic arm-assisted system and gap-balancing instrument in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 32, 685–692. [CrossRef] [PubMed] [Google Scholar]
  5. Koutserimpas C, Saffarini M, Bonnin M, et al. (2024) Optimizing the patellofemoral compartment in total knee arthroplasty: Is it time for dynamic assessment? Knee Surg Sports Traumatol Arthrosc 33, 387–392. https://doi.org/10.1002/ksa.12450. [Google Scholar]
  6. Lee JH, Kwon SC, Hwang JH, et al. (2024) Functional alignment maximises advantages of robotic arm-assisted total knee arthroplasty with better patient-reported outcomes compared to mechanical alignment. Knee Surg Sports Traumatol Arthrosc 32, 896–906. [CrossRef] [PubMed] [Google Scholar]
  7. Choi BS, Kim SE, Yang M, et al. (2023) Functional alignment with robotic-arm assisted total knee arthroplasty demonstrated better patient-reported outcomes than mechanical alignment with manual total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 31, 1072–1080. [CrossRef] [PubMed] [Google Scholar]
  8. Gregori P, Koutserimpas C, De Fazio A, et al. (2025) Functional knee positioning in patients with valgus deformity undergoing image-based robotic total knee arthroplasty: Surgical technique. SICOT J 11, 7. [PubMed] [Google Scholar]
  9. Shatrov J, Battelier C, Sappey-Marinier E, et al. (2022) Functional alignment philosophy in total knee arthroplasty – Rationale and technique for the varus morphotype using a CT based robotic platform and individualized planning. SICOT J 8, 11. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
  10. Batailler C, Greiner S, Rekik H-L, et al. (2024) Intraoperative patellar tracking assessment during image-based robotic-assisted total knee arthroplasty: technical note and reliability study. SICOT J 10, 44. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
  11. Oussedik S, Abdel MP, Victor J, et al. (2020) Alignment in total knee arthroplasty. Bone Joint J 102-B, 276–279. [CrossRef] [PubMed] [Google Scholar]
  12. Gregori P, Koutserimpas C, Giovanoulis V, et al. (2025) Functional alignment in robotic-assisted total knee arthroplasty for valgus deformity achieves safe coronal alignment and excellent short-term outcomes. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1002/ksa.12585. [Google Scholar]
  13. Jeffrey M, Marchand P, Kouyoumdjian P, Coulomb R (2024) Short-term functional outcomes of robotic-assisted TKA are better with functional alignment compared to adjusted mechanical alignment. SICOT J 10, 2. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
  14. Koutserimpas C, Garibaldi R, Olivier F, et al. (2025) Tibial implant varus >3° does not adversely affect outcomes or revision rates in functionally aligned image-based robotic total knee arthroplasty in a minimum of 2-year follow-up. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1002/ksa.12659. [PubMed] [Google Scholar]
  15. Koutserimpas C, Gregori P, Andriollo L, et al. (2025) Comparable outcomes between cruciate-substituting and posterior-stabilized inserts in robotic total knee arthroplasty under the functional alignment principles. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1002/ksa.12654. [Google Scholar]

Cite this article as: Koutserimpas C, Andriollo L, Gregori P, Veizi E, Tandogan R, Lustig S & Dretakis K (2025) Total knee arthroplasty after anterior cruciate ligament reconstruction with the use of image-based robotic technology and functional alignment. SICOT-J 11, 30. https://doi.org/10.1051/sicotj/2025025.

All Tables

Table 1

The demographics of the studied cohort are presented. Data are reported as mean values with standard deviations (SD).

Table 2

The preoperative and postoperative evaluation with the Knee Society Score (KSS) is presented. Data are reported as mean values with standard deviations (SD). Independent t-test was used for the comparison of the mean values. Statistical significance was defined as P < 0.05. In italics the statistically significant values.

Table 3

Advantages and disadvantages of the use of the image-based robotic system in total knee arthroplasty in patients with previous anterior cruciate ligament reconstruction.

All Figures

thumbnail Figure 1

Intraoperative identification of the tibial fixation staple using the image-based robotic system. The robotic probe is utilized to locate the fixation staple, which was entirely covered by bone and not visible through direct visualization. The system provides real-time spatial feedback, displaying the distance to the bone and mapping reference points for accurate positioning. The inset image on the right shows the corresponding CT scan with the staple’s location, ensuring precise intraoperative identification.

In the text
thumbnail Figure 2

Intraoperative verification of tibial component rotation using robotic guidance. The rotation of the tibial implant is assessed with reference to Akagi’s line. The use of the probe ensures accurate rotational positioning before final implantation.

In the text
thumbnail Figure 3

Preoperative planning of the femoral component and boxing cut using an image-based robotic system. CT-based imaging was utilized to precisely position the femoral implant while ensuring the cutting box aligned with the head of the femoral fixation screw. The robotic system provided detailed visualizations in transverse, coronal, and sagittal planes.

In the text
thumbnail Figure 4

Intraoperative verification of the mediolateral positioning of the femoral implant using a robotic-assisted system. The robotic probe was utilized to confirm the precise mediolateral placement of the femoral component before proceeding with the box cut. This step ensured accurate localization of the femoral fixation screw head, preventing blind bone removal and allowing for a controlled resection while preserving bone stock.

In the text

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.