Table 2

Biomechanical test series with TIFI.

Authors Test models Type of sacral fracture Vertical load Summary/Conclusion
Dienstknecht et al. [29] 6 freshly frozen human pelvis 6 unilateral transforaminal 3 cycles of 70% of the former body mass were performed max load not reported
  • TIFI compared to two IS screws were tested

  • TIFI was not inferior compared to IS screws

Salasek et al. [30] Computed image finite element Unilateral transforaminal 250–500 N
  • TIFI has enough stiffness for unstable transforaminal fractures

  • TIFI is superior compared to two IS screws

Shinohara et al. [31] 8 synthetic pelvic models 8 unilateral transforaminal
  • Max. load 1057 N

  • Mean load for a 5 mm gap: 438 N

  • Double (dual) TIFI with crosslinks were used

  • TIFI was superior compared to plate

Vigdorchik et al. [32] 6 synthetic pelvic models
  • 3 unilateral transforaminal

  • 3 IS joint dislocation

  • 5 cycles; static load 10 mm/sec

  • Mean load for a 7 mm gap: 81 N (sacral fracture)

  • Mean load for a 7 mm gap: 71 N (IS dislocation)

  • Sacral fracture: TIFI was inferior compared to TIFI plus IS screw

  • Sacral fracture: TIFI was equal to IS screw alone

  • IS dislocation: TIFI was the weakest construct

Chaiyamongkol et al. [33] 12 synthetic pelvic models 12 unilateral transforaminal
  • 280, 161, and 510 N

  • Static load 10 mm/min

  • Mean load for a 5 mm gap: 287 N

  • TIFI without IS screw was not tested

  • TIFI plus one IS screw better than two IS screws or plate