Open Access

Table 2

Results of the traction rupture test.

Implant DSP Metallic IFS MILAGRO p-value
Maximum load (N) 98.8 ± 45.3 92.1 ± 32.4 109.4 ± 35.9 0.60
Stiffness (N/mm) 15.2 ± 7.0 13.9 ± 9.2 15.9 ± 15.2 0.92
Elongation at failure (mm) 36.0 ± 16.6 25.8 ± 8.3 32.1 ± 13.2 0.23
Elongation at yield (mm) 1.9 ± 1.2 2.96 ± 1.6 3.95 ± 2.4 0.05

DSP: double spike plate (Meira Corp, Nagoya, Aichi, Japan); metallic IFS: interference screw (CANNU-FLEX SILK SCREWS, Smith & Nephew sports medicine, Andover, MA).