Introduction

Dissociation of the polyethylene liner from the acetabular shell is a rare but catastrophic complication of total hip arthroplasty (THA). This has been well researched and it appears that certain implant designs have shown a greater risk of polyethylene liner dissociation (PLD). Despite the excellent track record with regard to the functional outcome and longevity, Pinnacle (DePuy), Harris–Galante (Zimmer) and Trident (Stryker) have been associated with PLD [15].

To the best of our knowledge, this is the first case where a PLD of an R3 acetabular component (Smith & Nephew) is being reported. This is a case report of a complex THA looking at implant choice in such cases in view of the recent literature.

Case report

A 56-year-old patient had a complex primary THA following failure of fixation of an intertrochanteric fracture treated with a Gamma nail, 7 years ago (Figure 1). His medical history included hypertension, heavy smoking, recreational drugs and excessive alcohol use.

thumbnail Figure 1

Painful right hip with collapse of the femoral head due to avascular necrosis.

THA was performed after removal of the Gamma nail with extension of the old incision using a posterolateral approach. The patient required release of the gluteus maximus, iliopsoas and anterior capsule to address the flexion contractures. An uncemented THA was performed using the Smith & Nephew R3 size 60 acetabular shell with 60/36 highly cross-linked polyethylene liner in the templated position. As the old fracture was extending up to the subtrochanteric region, a size 15 Echelon uncemented stem was used to bypass the subtrochanteric region. A CoCr 36 + 10 femoral head was used and was found to be stable (Figure 2A and B). Clinical and radiographical follow up of his THA was unremarkable for 5 years postoperatively. However, 5 years after the primary operation, the patient had a fall and presented with a new onset of hip pain, and radiographs showed radiolucency medial to the femoral neck in association with an eccentrically placed femoral head lying in contact with the acetabular metal shell. Careful evaluation of the radiographs showed that this radiolucency was consistent with a PLD (Figure 3A and B).

thumbnail Figure 2

A and B: Complex primary THA.

thumbnail Figure 3

A and B: Proximal migration of the femoral head and radiolucent ring suggestive of polyethylene liner dissociation.

Infection work up including inflammatory markers and a joint aspiration were clear; hence a single-stage revision was proposed. Patient declined surgery and opted to carry on for further 3 months at which point his pain worsened and hence revision arthroplasty was performed. Intraoperatively, the acetabular liner was found to be dissociated (Figure 4). The femoral component was stable and therefore retained. Surgery involved replacement of the acetabular shell and liner and the femoral head along with excision of a pseudo tumour resulting from metallosis. Trial reduction and range of movement testing were satisfactory, and hence the definitive components implanted included a Zimmer Continuum shell of size 64 mm with an offset liner and 36 mm + 12 Oxinium femoral head (Figure 5).

thumbnail Figure 4

Intra operative picture showing the dissociated liner.

thumbnail Figure 5

Radiographs after revision of the acetabular components.

The patient made an unremarkable recovery in the immediate postoperative period but presented with a dislocation of his THA within a month of his operation. He underwent closed reduction under general anaesthesia but unfortunately, continued to have persistent instability, and hence a month later the patient underwent a second revision. Intraoperatively, the components were reassessed and it was confirmed that both the acetabular shell and femoral stem were well fixed in excellent position and therefore both components were retained. The polyethylene was revised and a constrained liner was used instead with a new size of 36 mm + 12 Oxinium femoral head (Figure 6A and B).

thumbnail Figure 6

A and B: Constrained liner following the 2nd revision THA.

Discussion

PLD has been described in a number of studies in the past [17]. The dissociation can occur either during the early or late postoperative period after a THA. Dissociations occurring in the early postoperative period are likely to be the result of inadequate seating of the polyethylene liner and failure of the locking mechanism of the acetabular component [1, 2]. Late dissociations are believed to be either from femoral neck impingement against the polyethylene liner or edge loading resulting in fatigue failure of the locking mechanism [1, 3].

The radiolucency of the dissociated liner seen on plain radiographs has been described as the “crescent” sign while using ultrasound this has been described as the “tram track” sign [810]. PLD has been reported with some of the most commonly used implants [1, 2, 4, 6, 11]. However, to date there have been no reports of this complication with the Smith & Nephew R3 system.

This is a case of complex arthroplasty in a patient with compliance issues due to alcoholism and substance abuse. Fixation of the femoral neck fracture failed due to avascular necrosis of the femoral head which was revised to a THA using unconstrained implants as the patient was young and this was his primary arthroplasty. A large femoral head of size 36 mm was used for better stability and the patient did well for 5 years until he had the fall which resulted in the liner dissociation. When the patient presented with the PLD, the acetabular components were revised without using any constrained options. The patient presented with a dislocated hip within a month of having the first revision for which he had a closed reduction. As he continued to have instability, this was then revised using a constrained liner, and at 1-year follow up there has been no further dislocation or any other complications.

In hindsight, we feel that considering both the patient factors and abnormal anatomy, a more constrained prosthesis might have been a better implant of choice after the patient presented with the liner dissociation. The recent literature suggests excellent survivorship with dual mobility cups and has recommended them as the primary implant of choice in high-risk patients [11, 12]. Modern dual mobility designs have resolved complications such as intra-prosthetic dislocations which were reported with the first generation [13]. Dual mobility has the advantage of addressing the head neck ratio, jump distance and the requirement for a constrained liner all at once [11]. The literature suggests that dual mobility outperforms large diameter femoral heads and constrained liners at 10 years of follow up [14].

In summary, we recommend the use of dual mobility or constrained liners as the implants of choice in complex THA cases with compromised anatomy, previous dislocation and non-compliant patients to minimise the risk of further instability.

Conflict of interest

Authors certify that they have no financial conflict of interest.

References

  1. Grazette AJ, Foote J, Whitehouse MR, Blom AW (2016) A review of outcomes and modes of presentation following liner dissociation from Harris-Galante uncemented acetabular components. Hip Int 26, 149–152. [CrossRef] [PubMed] [Google Scholar]
  2. Langdown AJ, Pickard RJ, Hobbs CM, Clarke HJ, Dalton DJ, Grover ML (2007) Incomplete seating of the liner with the Trident acetabular system: a cause for concern? J Bone Joint Surg Br 89, 291–295. [CrossRef] [PubMed] [Google Scholar]
  3. Napier RJ, Diamond O, O’Neill CKJ, O’Brien S, Beverland DE (2017) The incidence of dissociated liners in 4,751 consecutive total hip arthroplasties using Pinnacle polyethylene acetabular liners. Hip Int 27, 537–545. [CrossRef] [PubMed] [Google Scholar]
  4. Kagan R, Anderson MB, Peters C, Pelt C, Gililland J (2018) Pinnacle polyethylene liner dissociation: a report of 3 cases. Arthroplast Today 4, 441–446. [CrossRef] [PubMed] [Google Scholar]
  5. Yun A, Koli EN, Moreland J, et al. (2016) Polyethylene liner dissociation is a complication of the DePuy Pinnacle Cup: a report of 23 cases. Clin Orthop Relat Res 474, 441–446. [CrossRef] [PubMed] [Google Scholar]
  6. Neumann D, Dorn U (2003) Polyethylene liner dislocation in Harris Galante acetabular components]. Z Orthop Ihre Grenzgeb 141, 190–194. [Google Scholar]
  7. Saito S, Ryu J, Seki M, Ishii T, Saigo K, Mori S (2008) Analysis and results of dissociation of the polyethylene liner in the Harris-Galante I acetabular component. J Arthroplasty 23, 522–526. [CrossRef] [PubMed] [Google Scholar]
  8. Jang JH, Jin W, Chun YS, et al. (2014) The tram track sign: a characteristic sonographic feature of polyethylene liner dissociation after total-hip arthroplasty. J Ultrasound Med 33, 1931–1937. [CrossRef] [PubMed] [Google Scholar]
  9. Jin W, Chun YS, Park SY, et al. (2009) Sonographic detection of acetabular polyethylene liner dissociation in total hip arthroplasty. J Ultrasound Med 28, 1597–1600. [CrossRef] [PubMed] [Google Scholar]
  10. White SP, Blom AW, Lee M, Smith EJ (2005) The crescent sign: dissociation of the polyethylene liner from a modular acetabular component in total hip arthroplasty. Skeletal Radiol 34, 620–624. [CrossRef] [PubMed] [Google Scholar]
  11. Darrith B, Courtney PM, Della Valle CJ (2018) Outcomes of dual mobility components in total hip arthroplasty: a systematic review of the literature. Bone Joint J 100-B, 11–19. [Google Scholar]
  12. Jones CW, De Martino I, D’Apolito R, Nocon AA, Sculco PK, Sculco TP (2019) The use of dual-mobility bearings in patients at high risk of dislocation. Bone Joint J. 101, 41–45. [Google Scholar]
  13. Neri T, Philippot R, Klasan A, et al. (2018) Dual mobility acetabular cups for total hip arthroplasty: advantages and drawbacks. Expert Rev Med Dev 15, 835–845. [CrossRef] [PubMed] [Google Scholar]
  14. Laurendon L, Philippot R, Neri T, Boyer B, Farizon F (2018) Ten-year clinical and radiological outcomes of 100 total hip arthroplasty cases with a modern cementless dual mobility cup. Surg Technol Int 32, 331–336. [PubMed] [Google Scholar]

Cite this article as: Parkar AAH, Sukeik M, El-Bakoury A & Powell J (2019) Acetabular liner dissociation: A case report and review of the literature. SICOT-J, 5, 31

All Figures

thumbnail Figure 1

Painful right hip with collapse of the femoral head due to avascular necrosis.

In the text
thumbnail Figure 2

A and B: Complex primary THA.

In the text
thumbnail Figure 3

A and B: Proximal migration of the femoral head and radiolucent ring suggestive of polyethylene liner dissociation.

In the text
thumbnail Figure 4

Intra operative picture showing the dissociated liner.

In the text
thumbnail Figure 5

Radiographs after revision of the acetabular components.

In the text
thumbnail Figure 6

A and B: Constrained liner following the 2nd revision THA.

In the text

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.