Table 1
Biomechanical comparison of Femoral fixation methods.
Femoral fixation | Premature rupture during cycling | Ultimate load (N) (Mean ± SD) | Elongation (mm) (Mean ± SD) | Stiffness (N/mm) (Mean ± SD) |
---|---|---|---|---|
Group F1: Biosure© RG 5 mm 5 mm | 5/5 (100%) | |||
Group F2: Biosure© RG 7 mm | 2/5 (40%) | 255 ± 170 | 6.7 ± 4.5 | 90 ± 5 |
Group F3: Healicoil Regenesorb 4.75 mm Anchor © | 3/5 (60%) | 308 ± 44 | 7.4 ± 1.4 | 119 ± 20 |
Group F4: Ultrabutton© | 2/5 (40%) | 427 ± 215 | 10.7 ± 3.8 | 111 ± 21 |
P-value | 0.506 | 0.389 | 0.208 |