Introduction

The arthroscopic repair of a massive rotator cuff tear could be surgically challenging with a high incidence of retear [13], and complete repair is complicated because of tendon retraction, muscle atrophy, and fatty degeneration [4]. The suture technique for grasping the torn tendon should provide high initial fixation strength, to prevent gap formation and maintain mechanical stability until healing has occurred [5, 6]. The Mason-Allen stitch (MAS) is known to be the strongest suture configuration [7]. Habermeyer et al. modified this configuration to be done arthroscopically, known as the modified Mason-Allen stitch (MMAS) [8]. A continuous argument regarding the best surgical technique and suture configuration to treat massive rotator cuff tears remains a surgical challenge [8]. Suture cutting out of the tendon and retear medial to tendon muscle junction are frequently reported issues that often require complex surgical techniques and additional anchors, which can be disadvantageous for surgeons and increase the economic burden for the patient [911]. Despite recent advancements in arthroscopic rotator cuff repair, the outcomes remain unsatisfactory, with no consensus reached [11, 12]. Therefore, innovations in this respect are necessary [11]. So far, limited studies have shown the comparative efficacy of different suture techniques [11]. Accordingly, surgeons base their choices on personal experience [11]. We described the double Mason-Allen configuration; its strength has not been previously reported. The primary purpose of our study is to measure the load to failure of the newly proposed double Mason-Allen suture configuration. The secondary purpose is to compare it to the commonly used suture configurations. We hypothesize that the double Mason-Allen repair configuration has the highest load to failure.

Materials and methods

This study was conducted on twenty-five male sheep infraspinatus tendons aged between 13 and 24 months. The specimens were obtained as whole sheep shoulders and used fresh without freezing. The soft tissue was dissected off the shoulder bony skeleton to expose the infraspinatus tendon and muscle. The infraspinatus tendon was thoroughly examined for any abnormalities or tears, then the tendon was cut from its humeral insertion, mimicking a full-thickness rotator cuff tear.

The specimens were randomly allocated to one of five groups (5 shoulders per group) in the following configurations:

Group 1 was allocated to the simple suture configuration (the suture tape passing only once from the articular side to the bursal side) [13].

Group 2 was allocated to the horizontal Mattress suture (the two limbs of the suture tape passed from the articular side to the bursal side) [13].

Group 3 was allocated to MMAS described by Habermeyer 2003 (two suture tapes, one in mattress fashion 1 cm from the torn tendon edge, and the second was passed in a simple fashion medial and between the mattress sutures) [6].

Group 4 was allocated to MAS described by Gerber et al. 1994 (one suture tape passed three times in the tendon substance, the first passage was from the articular to the bursal side 1 cm from the torn tendon edge, the second passage was from the bursal to the articular side about 1 cm from the first, finally, the last passage was from the articular to the bursal side between the first two passages and medial to them) [13].

Group 5 was allocated to double MAS suture (the MAS configuration was done twice on the same tendon specimen, where the 2 sutures were placed at a 5 mm distance from each other) (Figure 1).

thumbnail Figure 1

A diagrammatic representation of the double Mason-Allen suture configuration. A, B and C reveal different diagrammatic representations of the double Mason-Allen suture.

All the suture passages were done by Arthrex Scorpion® device using 1.3 mm suture tape (Arthrex, Naples, FL, USA) through the torn tendons.

Every specimen was then fixed to the test apparatus through its scapula. The suture tape-free end was hung to Sartorius® precision balance model QS64B (Sartorius AG, Germany). Reference Line Weights according to R111-1 OIML Class M1 was used to increase the weight on the suture tape in a controlled manner of 100 g until failure. Weight and load to failure were recorded for analysis for each specimen.

Statistical analysis of data

Collected data were stored in Microsoft® Excel® for Microsoft 365 MSO (Version 2302 Build 16.0.16130.20848) 64-bit and analyzed using the statistical package for social sciences, version 20.0 (SPSS Inc., Chicago, IL, USA). Quantitative data were expressed as mean ± standard deviation (SD). The ANOVA test was conducted, followed by post hoc analysis to determine differences between the groups in relation to load-to-failure responses. The confidence interval was set to 95% and the margin of error accepted was set to 5%. So, the P value was considered significant as the following: Probability (P value)

  1. P value less than 0.05 was considered significant.

  2. P value less than 0.001 was considered highly significant.

  3. P value greater than 0.05 was considered insignificant.

Results

Comparing the weights and loads to failure of the five suture configurations, the double Mason-Allen stitch had the highest mean load to failure of 423.30 ± 23.05 (Newtons) followed by the Modified Mason-Allen stitch. The difference between the two was statistically significant with a p-value < 0.05.

The results indicate that the proposed double Mason-Allen configuration has a higher rupture force by about 265% compared to the MMAS, which is a significant difference. Also, the percentage increase in rupture force for the double Mason-Allen was higher than all the other tested suture configurations.

The mean load causing suture failure between the five groups was statistically different, with a p-value < 0.001. The five groups can be arranged in decreasing order of mean load to failure as follows: double Mason-Allen, MMAS, MAS, horizontal mattress, and simple suture respectively (Table 1). Table 2 summarizes the pros and cons of each technique.

Table 1

Results obtained for the required load to rupture the tested suture configuration, including the new proposed double Mason-Allen.

Table 2

Summary of the pros and cons of the tested suture configurations.

Discussion

The optimal surgical technique and suture configuration for the arthroscopic repair of a massive rotator cuff tear is still a topic of debate [8]. This study demonstrated that the double Mason-Allen suture configuration had the highest load to failure, with a mean of 423.30 ± 23.05(Newtons). Conversely, the simple suture was the weakest. This study established that both the double Mason-Allen and modified Mason-Allen sutures had the highest biomechanical properties for the load to failure, suggesting the double MAS to be a better choice for repair configuration. To the best of the author’s knowledge, no studies in literature assessed the double Mason-Allen and compared it to the well-established suturing techniques.

The study had some limitations, such as not investigating other suture configurations that could have possibly shown more strength. Additionally, the challenge of osteoporotic bone and suture anchor pullout could not be tested as the bony insertion was removed. Lastly, fresh tendons were used for repair, which may not accurately replicate the tendons in chronic tears.

It has been demonstrated that the infraspinatus tendon of sheep has characteristics similar to those of the supraspinatus tendon in the human shoulder [14]. This makes it an effective model for studying rotator cuff diseases [14]. Repairing of chronic, massive rotator cuff tears is challenging and has a high rate of re-ruptures [15]. The weakest point of the repair is where the suture meets the tendon. So, achieving a secure tendon fixation is critical until biological healing occurs [15]. Various grasping techniques have been proposed to minimize suture failure and prevent cutting through the tendon [16]. In 2006, Tao and Kaltenbach showed the reproducibility and advantages of the modified Mason-Allen suture configuration and stated that it may improve the healing rates of rotator cuff repair done using a suture anchor [17]. In accordance with our results, in a study on fourteen sheep shoulders, Sileo and colleagues found that the modified Mason-Allen stitch and Mason-Allen stitch yield similar biomechanical features when suture-anchored into bone [18]. In an experimental in-vitro analysis in 2004 Ma and colleagues concluded that the MMAS provides strength with a mean value of 233 ± 40 (Newtons), comparable with that of the MAS with a mean value of 246 ± 40 (Newtons) [19].

For massive, contracted tears, it is often only possible to achieve single-row fixation [16]. Using multiple pairs of Mason-Allen stitches creates additional fixation points, resulting in reduced load per point and increased load to failure that the soft tissue can bear [20].

In our study, suture failure in all the 25 studied specimens occurred at the tendon grip interface, and no suture material tear was noted, this is explained by the very high resistance of the suture material used (suture tape) [21].

Conclusion

The double Mason-Allen repair configuration has the highest load to failure compared to the other known suture configuration to repair rotator cuff tears.

Recommendations

We recommend conducting a large-scale in vitro study using the double Mason-Allen suture as a preliminary step before performing surgical cuff repair. Subsequently, the short- and long-term outcomes of the suture on patients should be tested.

Funding

This research did not receive any specific funding.

Conflicts of interest

The authors declare they have no relevant financial or non-financial interests to report.

Data availability statement

The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Author contribution statement

Fourth author: Conceptualization, Methodology; first author: Writing original draft; fourth and first: Visualization, Investigation; third author: Supervision; first and second: Writing, Reviewing, and Editing.

Ethics approval

This study received ethical approval from the Ethics Committee of Ain Shams University under protocol number R 187/2022.

References

  1. Nagda S, Wiesel B, Abboud J, Salamone A, Sheth N, Foran J, Garstka J (2015) Patient perception of physician reimbursement in elective shoulder surgery. J Shoulder Elbow Surg 24(1), 106–110. [CrossRef] [PubMed] [Google Scholar]
  2. Okubo A, Yotsumoto T, Watanabe N, Kajikawa T, Nakajima S, Oshima Y, Iizawa N, Majima T (2021) Comparison of three suture-bridge techniques for large or massive rotator cuff tear with delamination. SICOT J 7, 41. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
  3. Roukoz S, George S, Aoun M, Daher M (2023) Combined latissimus dorsi transfer and deltoid flap for irreparable rotator cuff tear: a retrospective analysis. SICOT J 9, 35. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
  4. Galasso O, Riccelli DA, De Gori M, De Benedetto M, Orlando N, Gasparini G, Castricini R (2017) Quality of life and functional results of arthroscopic partial repair of irreparable rotator cuff tears. Arthroscopy 33(2), 261–268. [CrossRef] [PubMed] [Google Scholar]
  5. Scheibel MT, Habermeyer P (2003) A modified Mason-Allen technique for rotator cuff repair using suture anchors. Arthroscopy 19(3), 330–333. [CrossRef] [PubMed] [Google Scholar]
  6. Aydin N, Karaismailoglu B, Gurcan M, Ozsahin MK (2018) Arthroscopic double-row rotator cuff repair: a comprehensive review of the literature. SICOT J 4, 57. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
  7. Schneeberger AG, von Roll A, Kalberer F, Jacob HA, Gerber C (2002) Mechanical strength of arthroscopic rotator cuff repair techniques: an in vitro study. J Bone Joint Surg Am 84(12), 2152–2160. [CrossRef] [PubMed] [Google Scholar]
  8. Miyazaki A.N., Fregoneze M., Santos P.D., da Silva L.A., do Val Sella G., Zanella L.A., La Salvia J.C., Checchia S.L. (2014) Biomechanical evaluation between the modified Mason-Allen stitch and the locked double-tie stitch on the infraspinatus of sheep. Rev Bras Ortop 49(6), 630–635. [CrossRef] [PubMed] [Google Scholar]
  9. Castagna A, Garofalo R, Conti M, Borroni M, Snyder SJ (2007) Arthroscopic rotator cuff repair using a triple-loaded suture anchor and a modified Mason-Allen technique (Alex stitch). Arthroscopy 23(4), 440.e1–440.e4. [PubMed] [Google Scholar]
  10. OuYang J, Xiao X, Li B, Zeng L, Xiang Y, Qin Y, Jin T, Yang K (2024) A modified Mason-Allen suture enhancement technique (sunglasses loop) for single-row repair of medium-to-large rotator cuffs. Arthrosc Tech 13(7), 103007. [CrossRef] [PubMed] [Google Scholar]
  11. Wang W, Kang H, Li H, Li J, Meng Y, Li P (2021) Comparative efficacy of 5 suture configurations for arthroscopic rotator cuff tear repair: a network meta-analysis. J Orthop Surg Res 16(1), 714. [CrossRef] [PubMed] [Google Scholar]
  12. Roubion RC, Field LD (2023) “Double row rip-stop” arthroscopic repair technique for acute, retracted rotator cuff tears. Arthrosc Tech 12(9), e1523–e1527. [CrossRef] [PubMed] [Google Scholar]
  13. Gerber C, Schneeberger AG, Beck M, Schlegel U (1994) Mechanical strength of repairs of the rotator cuff. J Bone Joint Surg Br 76 (3), 371–380. [CrossRef] [Google Scholar]
  14. Gerber C, Schneeberger AG, Perren SM, Nyffeler RW (1999) Experimental rotator cuff repair: A preliminary study. J Bone Joint Surg Am 81(9), 1281–1290. [CrossRef] [PubMed] [Google Scholar]
  15. Lorbach O, Bachelier F, Vees J, Kohn D, Pape D (2008) Cyclic loading of rotator cuff reconstructions: single-row repair with modified suture configurations versus double-row repair. Am J Sports Med 36(8), 1504–1510. [CrossRef] [PubMed] [Google Scholar]
  16. Denard PJ, Burkhart SS (2011) Techniques for managing poor quality tissue and bone during arthroscopic rotator cuff repair. Arthroscopy 27(10), 1409–1421. [CrossRef] [PubMed] [Google Scholar]
  17. Tao SS, Kaltenbach J (2006) Arthroscopic placement of a modified Mason-Allen stitch. Arthroscopy 22(11), 1248.e1–1248.e3. [PubMed] [Google Scholar]
  18. Sileo MJ, Ruotolo CR, Nelson CO, Serra-Hsu F, Panchal AP (2007) A biomechanical comparison of the modified Mason-Allen stitch and massive cuff stitch in vitro. Arthroscopy 23(3), 235–240. [CrossRef] [PubMed] [Google Scholar]
  19. Ma CB, MacGillivray JD, Clabeaux J, Lee S, Otis JC (2004) Biomechanical evaluation of arthroscopic rotator cuff stitches. J Bone Joint Surg Am 86(6), 1211–1216. [CrossRef] [PubMed] [Google Scholar]
  20. Jost PW, Khair MM, Chen DX, Wright TM, Kelly AM, Rodeo SA (2012) Suture number determines strength of rotator cuff repair. J Bone Joint Surg Am. 94 (14) e100. [CrossRef] [PubMed] [Google Scholar]
  21. Borbas P, Fischer L, Ernstbrunner L, Hoch A, Bachmann E, Bouaicha S, Wieser K (2021) High-strength suture tapes are biomechanically stronger than high-strength sutures used in rotator cuff repair. Arthrosc Sports Med Rehabil 3 (3): e873–e880. [CrossRef] [PubMed] [Google Scholar]

Cite this article as: Haroun Y, Sobhy MH, Khater HA & Khater AH (2024) Comparison between four different suture configurations for rotator cuff repair: a biomechanical animal study. SICOT-J 10, 39

All Tables

Table 1

Results obtained for the required load to rupture the tested suture configuration, including the new proposed double Mason-Allen.

Table 2

Summary of the pros and cons of the tested suture configurations.

All Figures

thumbnail Figure 1

A diagrammatic representation of the double Mason-Allen suture configuration. A, B and C reveal different diagrammatic representations of the double Mason-Allen suture.

In the text

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.