Volume 10, 2024
Article Number 44
Number of page(s) 6
Section Knee
DOI https://doi.org/10.1051/sicotj/2024037
Published online 29 October 2024

© The Authors, published by EDP Sciences, 2024

Licence Creative CommonsThis is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Introduction

In total knee arthroplasty (TKA), both surgeon and patient expectations are continuously rising. The contemporary objectives of primary TKA include not only improved functional outcomes but also the capability to engage in more demanding daily activities, such as kneeling and recreational sports, which were previously unattainable [1]. As a common cause of discomfort, pain, or preclusion from these activities, almost 50% of these patients point to the anterior part of the knee [2, 3].

Restoration of the anterior knee compartment is being studied increasingly with the development of personalized surgery. Even minor imperfections in the restoration of this compartment can lead to patellar maltracking, severe pain, and poor functional outcomes [4]. During knee flexion, the patella’s movement is primarily guided by the retinaculum between 0 and 30° and subsequently by the shape of the trochlear groove during mid-flexion and high flexion [5, 6]. This dynamic relationship is influenced by various factors, including surgical techniques (particularly implant positioning, femoral, and patellar sizing) and implant characteristics (such as trochlear depth and shape, sagittal curvature, and patellar component design) [7, 8]. New robotic-assisted technologies now enable surgeons to address the patellofemoral and tibiofemoral compartments independently and to reconsider thinking about the anterior knee compartment. Indeed, thanks to the preoperative CT scan, the shape of the trochlea is materialized on the screen during the planning. The femoral implant can be superimposed and positioned precisely in the same position as the native trochlea if wanted. Consequently, recreating the physiologic position and depth of the trochlea groove appears easier with this robotic assistance.

However, the assessment of patellar tracking during surgery is still subjective and at the surgeon’s discretion. Image-based robotic-assisted systems can now visualize patellar tracking intraoperatively; however, no standardized measurement of this tracking has been established.

This study aimed 1) to describe the assessment of the patellar tracking during robotic-assisted total knee arthroplasty (TKA), 2) to explain a new measurement technique for evaluating the evolution of this patellar tracking, and 3) to assess its repeatability and reliability.

The hypothesis was that this measurement technique was reliable and consistently repeatable.

Material and methods

Study design

This retrospective study included 20 primary total knee arthroplasties performed using an image-guided robotic system in a varus population. Prior to surgery, each patient underwent a dedicated CT scan with 3D reconstructions.

Surgical technique

All surgeries were performed by an orthopaedic surgeon with more than five years of experience in robotic-assisted TKA and more than 200 cases per year.

The MAKO robotic platform planning software (MAKO, Stryker Corporation, Mahwah, NJ, USA) allowed preoperative implant planning using a patient-specific CT-based bone model and virtual implant templates. As shown previously, the 3D implant model followed the bone anatomy and was accurate to within 1 mm [9].

  • First assessment of patellar tracking (before bone cuts)

The patellar tracking is assessed before the bone resections. The patella is reduced in front of the femur. It’s essential to make a landmark on the anterior face of the patella to use the same landmark at each patellar tracking assessment (Video 1). This landmark, performed with the electric scalpel, is located in the middle of the patella in the mediolateral and proximal-distal axis. The sharp probe is used to perform the patellar tracking because it is more stable on the bone (Video 1). It is crucial not to constrain the patella with the probe to ensure the real patellar positioning in all the arcs of knee flexion. The knee goes from full extension and is flexed progressively (Video 2). The MAKO engineer records the patella positioning every 10° of flexion between the full extension and 90° knee flexion (Video 2, Figure 1).

thumbnail Figure 1

Patellar tracking before the bone resection on the CT scan of the knee planning.

The patella’s positioning can be assessed in the coronal (mediolateral) and sagittal (anteroposterior) positions.

  • TKA planning and bone cuts

The positioning of the implant followed the principles of functional alignment described previously [10]. If a lateral translation of the patella is identified before the planning during the patellar tracking, the femoral implant can be positioned a little more lateral and with higher lateral rotation. Bone cuts were then executed with the robotic arm.

  • New assessment of patellar tracking (after bone cuts)

The final step is the verification of the functional plan. Limb alignment and gap sizes can be assessed clinically and with real-time feedback from the robot, either with trial or definitive implants, or both. The patellar tracking is again evaluated to analyze the evolution of the patellar tracking.

In the case of patellar resurfacing, this assessment is performed before (after the bone resections) and after the patellar resurfacing. The evaluation before the patellar resurfacing allows us to adjust the patellar button positioning if needed.

It is essential to use the same patellar landmark with the sharp probe. The knee goes again from full extension and is flexed progressively without constraint on the patella. The MAKO engineer records the patella positioning every 10° of flexion between the full extension and 90° knee flexion, as before the bone cuts (Video 3).

The patella’s positioning can be compared to the preoperative positioning in the coronal (mediolateral positioning) and sagittal planes (anteroposterior positioning) (Figure 2).

thumbnail Figure 2

The new assessment of the patellar tracking after the bone resections and with the definitive implants. The two patellar tracking (before and after the TKA) can be compared on the same screening.

Measurement technique

All measurements were made using the MAKO robotic platform planning software (MAKO, Stryker Corporation, Mahwah, NJ, USA). A calibrated millimetre scale allowed accurate and reliable measurements with an accuracy of 1 mm.

For the measurements of the patellar tracking, we used a sagittal view (Figure 3a) and a coronal view (Figure 3b) of the knee on the MAKO software. On these two views, we measured the difference between the patellar tracking before the bone cuts and after the bone cuts with the definitive implants. On the sagittal view, we measured this anteroposterior difference every 10° between 0° and 90°, considering 0° as the vertical position and 90° as the horizontal position distally compared to the knee rotation centre (Figure 4a). On the coronal view, we measured this mediolateral difference in three positions corresponding to 30°, 70° and 90° (Figure 4b).

thumbnail Figure 3

(a) Sagittal and (b) coronal views of the knee with patellar tracking before and after the bone cut.

thumbnail Figure 4

(a) Sagittal view of the knee with the localization of the measurements at 0° and 90°. (b) Coronal view of the knee with the measurements localized at 30°, 70°, and 90°.

A difference inferior to 1 mm was considered null. Measurements were performed by two independent reviewers (an orthopaedic surgeon and a medical student) for all measurements to assess the reliability of each measurement. To determine intra-observer variability, the second observer measured the patients twice, four weeks apart. Both observers were trained on the MAKO platform to learn the measurement technique.

Statistical analysis

Statistical analysis was performed using the XL STAT software (Version 2021.2.1, Addinsoft Inc., Paris, France). An intraclass correlation coefficient evaluated the inter- and intra-observer reliabilities of the measurements. The strength of agreement for the kappa coefficient was interpreted as follows: <0.20 = unacceptable, 0.20–0.39 = questionable, 0.40–0.59 = good, 0.60–0.79 = very good, and 0.80–1 = excellent [11]

Results

The mean age was 68.7 years old ± 5.2 [61; 75], the mean body mass index was 28.8 kg/m2 ± 4.2 [21.4; 36.2], the mean HKA angle was 176.3° ± 3.7° [174.1.4; 179.7].

The values of the evolution of patellar tracking are summarized in Table 1. There was a tendency to obtain a patella closer to the femur at the proximal part of the femoral implant. All the mediolateral modifications were a decrease in the patellar lateralization.

Table 1

Values of the evolution of patellar tracking.

The radiographic measurements showed very good to excellent intra-observer and inter-observer agreements (Table 2).

Table 2

Intraobserver and inter-observer coefficients for the evolution of patellar tracking.

Discussion

The main finding of this study was the description of a new technique for intraoperative measurement of patellar tracking evolution during total knee arthroplasty. This technique has demonstrated both reliability and repeatability.

Using classic mechanical instrumentation, the surgeons usually focus on the tibiofemoral joint, overlooking the patellofemoral joint’s complexity. Surgeons typically manage this by adjusting the rotation and lateralization of the femoral component. This unawareness of the anterior compartment can lead to early complications or revisions due to patellofemoral problems [12]. The advantage of this new patellar tracking assessment technique is its intraoperative application, allowing for adjustments before the implantation of definitive components. Based on real-time assessment, surgeons can modify femoral and tibial implant positioning (e.g., lateral position, rotation) and the placement of the patellar button. In addition, this could provide new insight for decision-making regarding patella resurfacing or not and how patella cut should be performed (direction in 3D and amount of bone to be removed). Every modification can be assessed immediately, including its impact on patellar tracking. This could decrease the complications of the extensor mechanism after TKA, particularly the patellar instability.

Personalized alignment techniques, performed with an image-based robotic-assisted system, have demonstrated superior understanding and restoration of the trochlear groove compared to mechanical or kinematic alignment methods [10, 13]. Mechanical alignment often results in the trochlear groove being positioned farthest from its native anatomy, while kinematic alignment can lead to unsafe coronal implant positioning in approximately 13% of cases and internal rotation of the femoral component beyond 3° in over 25% of cases [10]. By contrast, the functional alignment had only 3.2% of patients outside coronal and 1.7% outside rotational safe zones [10]. Nevertheless, these personalized surgeries tend to understaff the trochlea and the anterior compartment, as reported in an in-vitro study using an image-based robotic system [10]. Recent studies showed the same tendency with a patellar tracking closer to the femoral implant between 0° and 50° [14, 15]. This study also showed significant patellar tracking modification in the sagittal plane during early flexion between 0° and 40° (3.2–4.2), likely due to proximal trochlear understuffing of the femoral implant. In the coronal plane, we also observed a reduction of the lateral translation for some patients with a preoperative lateral subluxation of the patella. Interestingly that this lateral translation can be assessed during surgery and corrected immediately.

Shatrov et al. described a similar analysis with an image-less robotic-assisted system [16]. They described the patella centre of rotation, with a high variability of modification after TKA in a small cohort. Further analyses are necessary to better understand the evolution of patellar tracking according to surgical planning.

This measurement technique has demonstrated its inter and intra-observer reliability and repeatability. The precision of measurement, facilitated by preoperative CT scans, was accurate to within 1 mm. The main difficulty in the measurement was the difference inferior to 1 mm because it cannot be assessed. However, a difference inferior to 1 mm can be considered insignificant clinically. Prior training with the robotic software was essential for observers’ accurate interpretation of measurements.

This study had several limitations. First, the MAKO system was necessary to measure the CT scan with the implants in place. Secondly, the measurements are based on CT scans and do not account for cartilage thickness. However, cartilage thickness did not impact the measurements since we evaluated differences between the two periods. The first measurements were difficult to perform, underscoring the importance of thorough training for observers using the MAKO software. This study did not evaluate the impact of these measurements on clinical outcomes; further studies are needed to explore these implications.

To our knowledge, this is the first study to describe a measurement technique for evaluating the evolution of patellar tracking after image-based robotic-assisted total knee arthroplasty. This study did not aim to interpret the evolution of patellar tracking but to develop a reliable measurement technique.

Conclusions

This new measurement technique evaluated the evolution of patellar tracking after total knee arthroplasty with good intra and inter-observer reliability. A further study should be conducted to correlate the development of patellar tracking after TKA with the functional results.

Funding

This research did not receive any specific funding.

Conflicts of interest

No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.

CB: Consultant for Smith and Nephew and Stryker.

SG, HLR: No conflict of interest.

FO: Employee for Stryker.

ES: Consultant for Smith and Nephew

SL: Consultant for Heraeus, Stryker, Depuy Synthes, Smith and Nephew. Institutional research support to Lepine and Amplitude. Editorial Board for Journal of Bone and Joint Surgery (Am).

Data availability statement

Data associated with this article cannot be disclosed due to legal reasons.

Author contribution statement

CB: Conceptualization, Methodology, Data curation, Writing original draft.

SG: Data curation, Writing original draft.

HLR: Data curation, Writing original draft.

FO: Validation, Data curation, Writing, Reviewing and Editing.

ES: Validation, Writing, Reviewing and Editing.

SL: Conceptualization, Supervision, Validation, Writing, Reviewing and Editing.

Ethics approval

Data collection and analysis were performed in accordance with the MR004 reference methodology of the Commission Nationale de l’Informatique et des Libertés (ref. 2229975V0), obtained on 6 May 2023. The study has been registered and deposited on the Health Data Hub website.

Informed consent

Written informed consent was obtained from all patients and/or families.

Supplementary materials

Video 1: Video showing the landmark for the patellar tracking in the middle of the patella and the sharp probe on the patella to track the patellar positioning during knee flexion. Access here

Video 2: Recording of the patellar positioning in extension and flexion before the bone cuts. Access here

Video 3: Recording of the patellar positioning in extension and flexion after the bone cuts with the trial implants. Access here

References

  1. Baker PN, Rushton S, Jameson SS, Reed M, Gregg P, Deehan DJ (2013) Patient satisfaction with total knee replacement cannot be predicted from pre-operative variables alone: A cohort study from the National Joint Registry for England and Wales. Bone Joint J, 95-B(10), 1359–1365. [CrossRef] [PubMed] [Google Scholar]
  2. Scuderi GR, Insall JN, Scott NW (1994) Patellofemoral pain after total knee arthroplasty. J Am Acad Orthop Surg 2(5), 239–246. [Google Scholar]
  3. Ranawat CS, White PB, West S, Ranawat AS (2017) Clinical and radiographic results of attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis. J Arthroplasty, 32(2), 431–436. [Google Scholar]
  4. Lustig S, Servien E, Batailler C (2023) How to optimize patellar tracking in knee arthroplasty? Orthop Traumatol Surg Res, 109(1S), 103458. [Google Scholar]
  5. Barrett D, Brivio A (2023) The third compartment of the knee: an update from diagnosis to treatment. EFORT Open Rev 8(5), 313–318. [Google Scholar]
  6. Iranpour F, Merican AM, Baena FR, Cobb JP, Amis AA (2010) Patellofemoral joint kinematics: the circular path of the patella around the trochlear axis. J Orthop Res 28(5), 589–594. [Google Scholar]
  7. Varadarajan KM, Rubash HE, Li G (2011) Are current total knee arthroplasty implants designed to restore normal trochlear groove anatomy? J Arthroplasty 26(2), 274–281. [Google Scholar]
  8. Breugem SJ, Haverkamp D (2014) Anterior knee pain after a total knee arthroplasty: What can cause this pain? World J Orthop 5(3), 163–170. [Google Scholar]
  9. Sires JD, Wilson CJ (2021) CT validation of intraoperative implant position and knee alignment as determined by the MAKO total knee arthroplasty system. J Knee Surg, 34(10), 1133–1137. [Google Scholar]
  10. Shatrov J, Batailler C, Sappey-Marinier E, Gunst S, Servien E, Lustig S (2022) Correction notice to: Functional alignment philosophy in total knee arthroplasty – rationale and technique for the varus morphotype using a CT based robotic platform and individualized planning. SICOT J 8, 18. [Google Scholar]
  11. Regier DA, Narrow WE, Clarke DE, et al. (2013) DSM-5 field trials in the United States and Canada, Part II: test-retest reliability of selected categorical diagnoses. Am J Psychiatry 170(1), 59–70. [CrossRef] [PubMed] [Google Scholar]
  12. Lachiewicz PF, Soileau ES (2006) Patella maltracking in posterior-stabilized total knee arthroplasty. Clin Orthop Relat Res 452, 155–158. [Google Scholar]
  13. Lustig S, Sappey-Marinier E, Fary C, Servien E, Parratte S, Batailler C (2021) Personalized alignment in total knee arthroplasty: current concepts. SICOT J 7, 19. [Google Scholar]
  14. Kafelov M, Farhat J, Servien E, Lustig S, Batailler C (2023) New measurement technique for restoration of the trochlear offset after image-based robotic-assisted total knee arthroplasty: a reliability study. SICOT J 9, 29. [Google Scholar]
  15. Kafelov M, Batailler C, Servien E, Lustig S (2024) Restoration of the anterior compartment after robotic total knee arthroplasty significantly improves functional outcome and range of motion at 1 year. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1002/ksa.12337. [Google Scholar]
  16. Shatrov J, Khasian M, Lording T, Monk AP, Parker D, Lustig S (2024) Robotic assessment of patella tracking in total knee arthroplasty. J ISAKOS 9(5), 100287. [Google Scholar]

Cite this article as: Batailler C, Greiner S, Rekik H-L, Olivier F, Servien E & Lustig S (2024) Intraoperative patellar tracking assessment during image-based robotic-assisted total knee arthroplasty: technical note and reliability study. SICOT-J 10, 44. https://doi.org/10.1051/sicotj/2024037.

All Tables

Table 1

Values of the evolution of patellar tracking.

Table 2

Intraobserver and inter-observer coefficients for the evolution of patellar tracking.

All Figures

thumbnail Figure 1

Patellar tracking before the bone resection on the CT scan of the knee planning.

In the text
thumbnail Figure 2

The new assessment of the patellar tracking after the bone resections and with the definitive implants. The two patellar tracking (before and after the TKA) can be compared on the same screening.

In the text
thumbnail Figure 3

(a) Sagittal and (b) coronal views of the knee with patellar tracking before and after the bone cut.

In the text
thumbnail Figure 4

(a) Sagittal view of the knee with the localization of the measurements at 0° and 90°. (b) Coronal view of the knee with the measurements localized at 30°, 70°, and 90°.

In the text