Original Article
Biomechanical evaluation of annulus fibrosus repair with scaffold and soft anchors in an ex vivo porcine model
1
OrthopeadicResearch Laboratory, Aarhus University Hospital,
Aarhus, Denmark
2
VU University Medical Center, Department of Physics and Medical Technology, Research Institute MOVE,
Amsterdam, The Netherlands
* Corresponding author: rickers@clin.au.dk
Received:
18
September
2017
Accepted:
17
April
2018
Introduction: Altered biomechanical properties, due to intervertebral disc (IVD) degeneration and missing nucleus fibrosus, could be thought as one of the reasons for the back pain many herniation patients experience after surgery. It has been suggested to repair annulus fibrosus (AF) to restore stability and allow nucleus pulposus (NP) replacement and furthermore prevent reherniation. The aim of this study was to evaluate a new method for closing a defect in AF for use in herniation surgery.
Methods: Our repair method combines a polycaprolactone (PCL) scaffold plugging herniation and soft anchors to secure the plug. Ex vivo biomechanical testing was carried out in nine porcine lumbar motion segments. Flexion–extension, lateral bending and rotation were repeated three times: first in healthy specimens, second with a full thickness circular defect applied, and third time with the specimens repaired. Finally push out tests were performed to check whether the plug would remain in.
Results: Tests showed that applying a defect to the AF increases the range of motion (ROM), neutral zone (NZ) and neutral zone stiffness (NZS). In flexion/extension it was found significant for ROM, NZ, and NZS. For lateral bending and rotation a significant increase in ROM occurred. After AF repair ROM, NZ and NZS were normalized. All plugs remained in the AF during push out test up until 4000 N, but NP was squeezed out through the pores of the scaffold.
Discussion: A defect in the AF changes the biomechanical properties in the motion segment, changes that point to instability. Repairing the defect with a PCL plug and soft anchors brought the biomechanical behavior back to native state. This concept is promising and might be a viable way to repair the IVD after surgery.
Key words: Annulus fibrosus / Scaffold / Intervertebral disc herniation / Soft anchors / Suture / Biomechanics
© The Authors, published by EDP Sciences, 2018
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.