Introduction

The prevalence of osteoarthritis (OA) in both North America and Europe is predicted to increase approximately 40% from 2005 to 2030 [1]. Total knee arthroplasty (TKA) and total hip arthroplasty (THA) remain the preferred treatment for end-stage knee and hip OA after failed non-surgical treatment. Joint replacement surgery reliably relieves pain, aids patients to return to near normal function and improves the health-related quality of life [2, 3]. THA is very successful and is associated with reproducible clinical outcomes with over 95% survivorship at 10-year follow-up and 80% survivorship at 25-year follow-up [2, 3].

An estimated increase in primary THA by 200% and TKA by as much as 673% is predicted from 2005 to 2030. Revision THA and TKA are projected to increase by 137% and 601%, respectively, between 2005 and 2030 [4]. Currently in the UK, 35% of patients under the age of 65 years are having hip and knee replacements and 12% of those are below the age of 55 years. There is very high probability of these patients requiring revision surgery in future [5]. The younger patient groups, especially males below the age of 55 years, have a 10-year survivorship and 16-year survivorship of 80% and 33%, respectively [6]. The patient expectations vary between higher demand younger patients and elderly patients. High expectations from TKA and THA pose a tough challenge to deliver functionally. TKA failure is multifactorial but more than 50% of early revisions are secondary to instability, malalignment or malposition, and failure of fixation [7], most of which result from poor surgical technique. THA failure is multifactorial. Early revisions at less than five years are due to implant malposition, dislocation and infection [8] which are dependent on the surgical technique. Good surgical technique and accurate implantation of the components are likely to improve the longevity and function of both TKA and THA. With such a huge revision rate predicted, it is imperative for orthopaedic surgeons to get it right the first time. Computer navigation provides evidence-based advantages to help surgeons achieve this goal.

Computer assisted surgery (CAS) has been around since early 1990s. The first in vitro CAS was performed in 1991 by Professor Nolte at Muller’s laboratory with the insertion of computer navigated pedicle screw into a saw bone vertebra [9]. This marked the beginning of a new era of innovation in the field of CAS in orthopaedics. Numerous novel ideas, refinement of processes and equipment led to the first image-free system for total knee replacement in 1997 by Picard and Saragaglia, who pioneered this technology that demonstrated promising future prospects [10]. Twenty years on, the technological advances in computer assisted orthopaedic surgery (CAOS) have consistently produced excellent outcomes in TKA, THA, knee osteotomies, spinal surgery and oncological surgery. New applications are being explored in anterior cruciate ligament reconstruction, ankle, shoulder and trauma surgery. We explore the role and evidence of CAOS in TKA and THA.

Material and methods

The objective criteria for surgeon dependent factors influencing TKA and THA are broken down into individual entities. These lead to reproducible improvement of the mechanical, clinical and functional outcome.

With TKA the surgeon dependent factors influencing outcome are divided into pre-operative, per-operative and post-operative factors.

  1. Pre-operative assessment of patient’s anatomy, deformity (static and dynamic) and function using clinical, radiological and objective function assessments. The dynamic deformity can be assessed with navigation using extra-cutaneous trackers in clinic or gait lab.

  2. Per-operative assessment and near normal reconstruction of patient’s anatomy, alignment, ligament balance, mechanics and kinematics.

  3. Post-operative clinical, radiological and functional outcome assessment.

The surgeon dependent factors in THA for optimal clinical, radiological and functional outcome are:

  1. Acetabular implant orientation and positioning of the centre in the intended place.

  2. Restoration of the patient’s hip offset and optimal leg length.

TKA

Pre-operative accurate static and dynamic detection of patient’s anatomy and deformity

Most of the existing conventional pre-operative methods of assessment are static which include radiographs (antero-posterior knee view, lateral view, skyline patella femoral joint view and long leg alignment view). The reliability of these radiographs in pre-operative planning is controversial [11, 12].

The patient specific instrumentation (PSI) has been described in the literature. The PSI technology requires pre-operative computerised tomography (CT) or magnetic resonance imaging (MRI) scan to enable the manufacturing of patient specific cutting jigs. PSI adds increased pressure on the radiology department, increases cost and, for the patient, requires additional hospital visits for the scan in addition to the radiation exposure of the CT scan. In some studies, PSI has produced less accurate correction of alignment of TKA components in comparison with traditional instrumentation [13]. The role of PSI in TKA is still unclear in the literature.

Computer navigation offers pre-operative dynamic assessment of deformity, alignment and kinematics. A study on normal individuals enabled dynamic assessment of knee laxity and femoro-tibial mechanical axis (FTMA) at 0° extension and 15° of knee flexion [14]. This has been described with a standardised technique using fibroelastic straps around femur and tibia stabilising extra-cutaneous markers that allow 6° of freedom of movement and a device producing 10 Nm torque to assess knee ligament laxity [14]. The potential role of navigation technology in the dynamic assessment of osteoarthritic knees in the outpatient clinic setting offers exciting opportunity in future to be able to set target values to reproduce during the surgery.

Per-operative assessment, guidance of bone cuts and near normal reconstruction of patient’s anatomy, alignment, ligament balance and mechanics

CAOS aids intra-operative dynamic accurate assessment of anatomy of individual patients differentiating subtle variations between various patients. Utilising standard anatomical landmarks, the computer designs a three-dimensional (3D) image of the anatomy [15]. CAOS provides dynamic and real-time assessment of femoral-tibial mechanical axis (FTMA), measurements of coronal and sagittal knee alignment offering a potential alternative to radiographs [16], tibial rotational profile [17] and knee kinematics [18].

A pre-implantation assessment of knee coronal plane deformity through the range of movement has been analysed and classified to assess kinematics [18]. This Deep’s classification describes four main types (1, 2, 3 and 4) with eight sub-groups (1A, 1B, 2A, 2B, 3, 4A, 4B and 4C); Table 1 [18].

Table 1.

Deep’s classification of knee deformity as it flexes from an extended position.

CAOS works with both gap balanced resection or measured bone resection. Live information of resection gaps, implant size and three-dimensional alignment of implants can be seen on a computer screen and allows bespoke changes for each patient (Figure 1) [15].

thumbnail Figure 1.

The implant size and positioning of the rotational position of femoral cutting jig.

Ligament release and soft tissue balancing are traditionally done based on deformity in extension and 90° flexion. Extensive collateral ligament releases are associated with post-operative haematoma, increased risk of infection and wound complications [19]. Extensive soft tissue release may increase hospital stay and may necessitate more constrained TKA implants. The dynamic change during flexion influences collateral releases. The dynamic assessment in CAOS achieves coronal deformity correction and soft tissue balance without needing extensive collateral soft tissue release [20]. Navigation allows post-implantation assessment of restoration of FTMA and kinematics (Figure 2) [18].

thumbnail Figure 2.

The change in coronal deformity (X axis) of knee femoro-tibial mechanical alignment angle, as the knee flexes (Y axis): first part showing before the surgery valgus deformity first increases and then decreases as the knee flexes (Type 4A in Deep’s classification) and second part showing the axis achieved after surgery of same patient (Neutral alignment).

Post-operative clinical, radiological and functional outcome assessment

CAOS has proven post-operative clinical and radiological improvement of alignment of components [21, 22]. Decreased post-operative embolism [23] and blood loss were noted in all patients including ones with high Body Mass Index (BMI) [24, 25]. In the Australian joint registry, at 11 years post navigation TKA, the revision rates are lower than TKA done with traditional instrumentation in patients under 55 years of age [26] (Figure 3).

thumbnail Figure 3.

Australian Joint registry report 2016.

Earlier lack of evidence on post-operative functional improvement has been one of the reasons CAOS did not gain so much popularity. Some studies found no difference in functional outcome between CAOS group and traditional group [27, 28]. Outcome assessment factors are gross and not powered enough to detect the statistical difference in most small studies. These studies are underpowered for detection of functional improvement. But now there is evidence in the literature including meta-analysis that CAOS along with improved anatomical accuracy leads to functional improvement too [21, 29, 30].

THA

Acetabular implant orientation and positioning of the centre in the intended place

Lewinnek et al. described the safe zone of acetabular cup insertion as anteversion 15° ± 10°; inclination 40° ± 10° [31]. Acetabular cup orientation is crucial to good performance and longevity of THA. Acetabular cup malposition outside of the Lewinnek’s safe zone increases the risk of complications like decreased range of movements, increased rates of impingement, polyethylene wear, acetabular migration, recurrent dislocations, pelvic osteolysis and early failure rates [3234]. Acetabular cup orientation based on surgeon’s visual assessment often results in an inaccurate placement [35]. The natural acetabular orientation in arthritic hips is quite variable and knowledge of this using navigation provides a better understanding of this orientation in individual patients [36], which offers an advantage for precise positioning of the acetabular component. Several studies have demonstrated convincingly that the acetabular orientation is more precise with navigated THA [37, 38], acetabular component was beyond the Lewinnek’s safe zone only in 8.63% of hips in the navigated group, compared with 28.4% in the conventional group [39].

Hip offset and leg length correction

Traditionally offset in THA has always been described in relation to femoral offset. Hip offset is a concept which factors in acetabular offset and hip centre of rotation (COR). A leg length/offset discrepancy of more than 5 mm correlates with non-physiological kinematics of gait [40]. Other detrimental outcomes include low back pain, nerve injury, patient dissatisfaction and increased litigations [41]. Navigated THA achieved 95.39% hip offset within 6 mm and 96.04% leg length restored within 6 mm of control contra-lateral side [42]. Navigated THA reduces malpositioning and facilitates the insertion of components in near normal orientation in comparison with the traditional THA [39, 42]. Patients are usually on lateral position and the most accepted landmarks are anterior superior iliac spines (ASIS) and pubic symphysis in anterior pelvic plane [31]. In obese patients, difficult palpation of these landmarks is an issue [43]. In several other studies, difficulty in registering anterior pelvic plane in obese patients is not an issue [44, 45].

CAOS offers excellent teaching and training opportunities with resultant reduction of learning curve associated with improvement in cognitive skills. The real-time feedback improves precision and accuracy in surgical techniques. It is an excellent teaching and training tool [46]. There is a role for simulation surgical training to accelerate the learning curve and to reduce complications.

Results

In summary, despite initial learning curve and increased operating time in initial surgeries, regular use of CAOS in TKA demonstrates increased accuracy of implant alignment leading to better knee function and improved longevity of implant in younger patients [26, 47, 48]. CAOS in THA aids precise acetabular cup placement with decrease in the number of outliers, better hip offset and leg length discrepancy [39, 42]. It leads to consistent reproducible results in the hands of both experienced navigated surgeon and trainee surgeon [49].

Discussion

The probability of finding new medical devices approved 10 years ago is less than devices approved five years ago. Fifty percent of devices approved for introduction to the orthopaedic market place are unavailable at 10 years. Only 2% of them were deemed to have safety problems [50]. CAOS has been embraced cautiously by orthopaedic surgeons across the world. There is about 30% usage in Germany [51], 28.6% usage in Australia [26] and sporadic usage in the UK, North America, Brazil, France and Asia [51]. The common arguments for delayed adoption of CAOS are increased operative duration, risk of a fracture at tracker pin sites, cost and in the past no difference in functional outcome. There is a learning curve, after 20 cases a novice navigation surgeon and experienced navigation assisted surgeon had similar operative times; however, all the surgical implantations were precise even during the learning curve [52]. In any surgery irrespective of navigation the initial few cases take longer time, which is the case for conventional joint replacements as well. Somehow the expectation from navigation technology has been that it should be equal or take even less time than conventional from the very start, even in the initial cases. Some experienced orthopaedic surgeons unlike their younger counterparts, less acquainted with computer games and technology, may not like the idea of computers during surgery and thus resist CAOS. A risk of fractures at tracker pin sites is reported, but the risk was as low as 0.13% (one in 777) in one study. The fracture occurred in an 82-year-old female patient subsequent to a direct fall onto the operated knee requiring an intramedullary nail fixation [53]. This problem has been substantially reduced by use of bicortical tracker attachment pins. In our institution with over 5000 navigated surgeries, no case of fracture was noted. The initial setup cost incurred for navigation system proves to be advantageous in long term. The number of operating instrumentation trays can be reduced with navigation. Numerous recent studies have described improved functional outcome in the CAOS group compared to the traditional instrumented group [29, 30, 47]. The revision burden from failed TKA and THA includes costlier implants, multiple prolonged hospital admissions, antibiotics, blood products, social services and costs. Any technique to help reduce this should be helpful. Now the navigation technology has been integrated into the robotic surgery. The available orthopaedic robots all use navigation technology to guide cut orientation and to detect movements during surgery. Only add the cutting end as well to make it robotic. We believe the evidence of CAOS technology in orthopaedics is well established over two decades and our successful adoption in routine practice confirms the role of computer navigation in lower limb arthroplasty. This will evolve to become a standard practice in the future in various forms like PSI, 3D printing, accelerometer-based navigation, infrared-based application or robotics.

Conflict of interest

Kamal Deep is a consultant for BBraun. He is the secretary of the CAOS International Society. The institution received research funding from Stryker, BBraun, Bayer, Mathys, Zimmer, Convatec, and BBT.

References

  1. Helmick CG, Felson DT, Lawrence RC, Gabriel S, Hirsch R, Kwoh CK et al. (2008) Estimates of the prevalence of arthritis and other rheumatic conditions in the United State. Part I. Arthritis Rheum 58(1), 15–25. [CrossRef] [PubMed] [Google Scholar]
  2. Charnley J (1972) The long-term results of low-friction arthroplasty of the hip performed as a primary intervention. J Bone Joint Surg Br 54(1), 61–76. [Google Scholar]
  3. Ethgen O, Bruyere O, Richy F, Dardennes C, Reginster J-Y (2004) Health-related quality of life in total hip and total knee arthroplasty. A qualitative and systematic review of the literature. J Bone Joint Surg Am 86–A(5), 963–974. [CrossRef] [PubMed] [Google Scholar]
  4. Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4), 780–785. [Google Scholar]
  5. Briggs TW (2012) Getting it right first time. Improving the quality of orthopaedic care within the National health service in England, https://www.boa.ac.uk/pro-practice/review-of-orthopaedics-in-england/. [Google Scholar]
  6. Malchau H, Herberts P, Eisler T, Garellick G, Soderman P (2002) The Swedish total hip replacement register. J Bone Joint Surg Am 84–A(Suppl), 2–20. [CrossRef] [Google Scholar]
  7. Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM (2002) Insall award paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res 404, 7–13. [CrossRef] [Google Scholar]
  8. Delaunay C, Hamadouche M, Girard J, Duhamel A (2013) What are the causes for failures of primary hip arthroplasties in France? Clin Orthop Relat Res 471(12), 3863–3869. [CrossRef] [PubMed] [Google Scholar]
  9. Nolte LP, Zamorano LJ, Jiang Z, Wang Q, Langlotz F, Berlemann U (1995) Image-guided insertion of transpedicular screws. A laboratory set-up. Spine (Phila Pa 1976) 20(4), 497–500. [CrossRef] [PubMed] [Google Scholar]
  10. Saragaglia D, Picard F, Chaussard C, Montbarbon E, Leitner F, Cinquin P (2001) Computer-assisted knee arthroplasty: comparison with a conventional procedure. Results of 50 cases in a prospective randomized study. Rev Chir Orthop Reparatrice Appar Mot 87(1), 18–28. [PubMed] [Google Scholar]
  11. Choi WC, Lee S, An JH, Kim D, Seong SC, Lee MC (2011) Plain radiograph fails to reflect the alignment and advantages of navigation in total knee arthroplasty. J Arthroplasty 26(5), 756–764. [CrossRef] [PubMed] [Google Scholar]
  12. Rauh MA, Boyle J, Mihalko WM, Phillips MJ, Bayers-Thering M, Krackow KA (2007) Reliability of measuring long-standing lower extremity radiographs. Orthopedics 30(4), 299–303. [PubMed] [Google Scholar]
  13. Thienpont E, Schwab PE, Fennema P (2014) A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Joint J 96-B(8), 1052–1061. [CrossRef] [PubMed] [Google Scholar]
  14. Deep K (2014) Collateral ligament laxity in knees. What is normal? Clin Orthop Relat Res 472(11), 3426–3431. [CrossRef] [PubMed] [Google Scholar]
  15. Picard F, Deep K, Jenny JY (2016) Current state of the art in total knee arthroplasty computer navigation. Knee Surgery, Sport Traumatol Arthrosc 24(11), 3565–3574. [CrossRef] [Google Scholar]
  16. Clarke JV, Riches PE, Picard F, Deakin AH (2012) Non-invasive computer-assisted measurement of knee alignment. Comput Aided Surg 17(1), 29–39. [CrossRef] [PubMed] [Google Scholar]
  17. Russell DF, Deakin AH, Fogg QA, Picard F (2014) Repeatability and accuracy of a non-invasive method of measuring internal and external rotation of the tibia. Knee Surg Sports Traumatol Arthrosc 22(8), 1771–1777. [CrossRef] [PubMed] [Google Scholar]
  18. Deep K, Picard F, Baines J (2016) Dynamic knee behaviour: does the knee deformity change as it is flexed – an assessment and classification with computer navigation. Knee Surgery, Sport Traumatol Arthrosc 24(11), 3575–3583. [CrossRef] [Google Scholar]
  19. Kumar PJ, Dorr LD (1997) Severe malalignment and soft-tissue imbalance in total knee arthroplasty. Am J Knee Surg 10(1), 36–41. [Google Scholar]
  20. Goudie S, Deep K (2014) Collateral soft tissue release in primary total knee replacement. Comput Aided Surg 19(1–3), 29–33. [CrossRef] [PubMed] [Google Scholar]
  21. Hetaimish BM, Khan MM, Simunovic N, Al-Harbi HH, Bhandari M, Zalzal PK (2012) Meta-analysis of navigation vs conventional total knee arthroplasty. J Arthroplasty 27(6), 1177–1182. [CrossRef] [PubMed] [Google Scholar]
  22. Lee D-H, Park J-H, Song D-I, Padhy D, Jeong W-K, Han S-B (2010) Accuracy of soft tissue balancing in TKA: comparison between navigation-assisted gap balancing and conventional measured resection. Knee Surg Sports Traumatol Arthrosc 18(3), 381–387. [CrossRef] [PubMed] [Google Scholar]
  23. Kalairajah Y, Cossey AJ, Verrall GM, Ludbrook G, Spriggins AJ (2006) Are systemic emboli reduced in computer-assisted knee surgery?: a prospective, randomised, clinical trial J Bone Joint Surg Br 88(2), 198–202. [CrossRef] [PubMed] [Google Scholar]
  24. McConnell J, Dillon J, Kinninmonth A, Sarungi M, Picard F (2012) Blood loss following total knee replacement is reduced when using computer-assisted versus standard methods. Acta Orthop Belg 78(1), 75–79. [PubMed] [Google Scholar]
  25. Millar NL, Deakin AH, Millar LL, Kinnimonth AWG, Picard F (2011) Blood loss following total knee replacement in the morbidly obese: effects of computer navigation. Knee 18(2), 108–112. [CrossRef] [PubMed] [Google Scholar]
  26. NJRRA (2016) Australian National Joint Registry Annual Report 2016, https://aoanjrr.sahmri.com/annual-reports-2016. [Google Scholar]
  27. Kim Y-H, Park J-W, Kim J-S (2012) Computer-navigated versus conventional total knee arthroplasty a prospective randomized trial. J Bone Joint Surg Am 94(22), 2017–2024. [CrossRef] [PubMed] [Google Scholar]
  28. Allen CL, Hooper GJ, Oram BJ, Wells JE (2014) Does computer-assisted total knee arthroplasty improve the overall component position and patient function? Int Orthop 38(2), 251–257. [CrossRef] [PubMed] [Google Scholar]
  29. Khakha RS, Chowdhry M, Sivaprakasam M, Kheiran A, Chauhan SK (2015) Radiological and functional outcomes in computer assisted total knee arthroplasty between consultants and trainees – a prospective randomized controlled trial. J Arthroplasty 30(8), 1344–1347. [CrossRef] [PubMed] [Google Scholar]
  30. Alcelik IA, Blomfield MI, Diana G, Gibbon AJ, Carrington N, Burr S (2016) A comparison of short-term outcomes of minimally invasive computer-assisted vs minimally invasive conventional instrumentation for primary total knee arthroplasty: a systematic review and meta-analysis. J Arthroplasty 31(2), 410–418. [CrossRef] [PubMed] [Google Scholar]
  31. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2), 217–220. [CrossRef] [PubMed] [Google Scholar]
  32. Kummer FJ, Shah S, Iyer S, DiCesare PE (1999) The effect of acetabular cup orientations on limiting hip rotation. J Arthroplasty 14(4), 509–513. [CrossRef] [PubMed] [Google Scholar]
  33. Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ (1998) Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty 13(5), 530–534. [CrossRef] [PubMed] [Google Scholar]
  34. McCollum DE, Gray WJ (1990) Dislocation after total hip arthroplasty. Causes and prevention. Clin Orthop Relat Res 261, 159–170. [Google Scholar]
  35. Rittmeister M, Callitsis C (2006) Factors influencing cup orientation in 500 consecutive total hip replacements. Clin Orthop Relat Res 445, 192–196. [PubMed] [Google Scholar]
  36. Goudie ST, Deakin AH, Deep K (2015) Natural acetabular orientation in arthritic hips. Bone Joint Res 4(1), 6–10. [CrossRef] [PubMed] [Google Scholar]
  37. Liu Z, Gao Y, Cai L (2015) Imageless navigation versus traditional method in total hip arthroplasty: a meta-analysis. Int J Surg 21, 122–127. [CrossRef] [Google Scholar]
  38. Verdier N, Billaud A, Masquefa T, Pallaro J, Fabre T, Tournier C (2016) EOS-based cup navigation: randomised controlled trial in 78 total hip arthroplasties. Orthop Traumatol Surg Res 102(4), 417–421. [CrossRef] [PubMed] [Google Scholar]
  39. Xu K, Li YM, Zhang HF, Wang CG, Xu YQ, Li ZJ (2014) Computer navigation in total hip arthroplasty: a meta-analysis of randomized controlled trials. Int J Surg 12(5), 528–533. [Google Scholar]
  40. Renkawitz T, Weber T, Dullien S, Woerner M, Dendorfer S, Grifka J et al. (2016) Leg length and offset differences above 5 mm after total hip arthroplasty are associated with altered gait kinematics. Gait Posture 49, 196–201. [CrossRef] [PubMed] [Google Scholar]
  41. Maloney WJ, Keeney JA (2004) Leg length discrepancy after total hip arthroplasty. J Arthroplasty 19(4 Suppl 1), 108–110. [CrossRef] [PubMed] [Google Scholar]
  42. Ellapparadja P, Mahajan V, Deakin AH, Deep K (2015) Reproduction of hip offset and leg length in navigated total hip arthroplasty: how accurate are we? J Arthroplasty 30(6), 1002–1007. [CrossRef] [PubMed] [Google Scholar]
  43. Lee YS, Yoon TR (2008) Error in acetabular socket alignment due to the thick anterior pelvic soft tissues. J Arthroplasty 23(5), 699–706. [CrossRef] [PubMed] [Google Scholar]
  44. Hohmann E, Bryant A, Tetsworth K (2012) Anterior pelvic soft tissue thickness influences acetabular cup positioning with imageless navigation. J Arthroplasty 27(6), 945–952. [CrossRef] [PubMed] [Google Scholar]
  45. Todkar M (2008) Obesity does not necessarily affect the accuracy of acetabular cup implantation in total hip replacement. Acta Orthop Belg 74(2), 206–209. [PubMed] [Google Scholar]
  46. Picard F, Moholkar K, Gregori A, Deep K, Kinninmonth A (2014) Role of computer assisted surgery (CAS) in training and outcomes. Orthop Trauma 28(5), 322–326. [CrossRef] [Google Scholar]
  47. Rebal BA, Babatunde OM, Lee JH, Geller JA, Patrick DAJ, Macaulay W (2014) Imageless computer navigation in total knee arthroplasty provides superior short term functional outcomes: a meta-analysis. J Arthroplasty 29(5), 938–944. [CrossRef] [PubMed] [Google Scholar]
  48. Choong PF, Dowsey MM, Stoney JD (2009) Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 24(4), 560–569. [CrossRef] [PubMed] [Google Scholar]
  49. Schnurr C, Eysel P, Konig DP (2011) Do residents perform TKAs using computer navigation as accurately as consultants? Orthopedics 34(3), 174. [CrossRef] [PubMed] [Google Scholar]
  50. Bhattacharyya T, Blyler C, Shenaq D (2006) The natural history of new orthopaedic devices. Clin Orthop Relat Res 451, 263–266. [CrossRef] [PubMed] [Google Scholar]
  51. Picard F, Clarke J, Deep K, Gregori A (2014) Computer assisted knee replacement surgery: Is the movement mainstream? Orthop Muscular Syst 3(2), 153. [CrossRef] [Google Scholar]
  52. Smith BRK, Deakin AH, Baines J, Picard F (2010) Computer navigated total knee arthroplasty: the learning curve. Comput Aided Surg 15(1–3), 40–48. [CrossRef] [PubMed] [Google Scholar]
  53. Harvie P, Sloan K, Beaver RJ (2011) Computer navigated total knee arthroplasty: aspects of a single unit’s experience of 777 cases. Comput Aided Surg 16(4), 188–195. [CrossRef] [PubMed] [Google Scholar]

Cite this article as: Deep K, Shankar S & Mahendra A (2017) Computer assisted navigation in total knee and hip arthroplasty. SICOT J, 3, 50

All Tables

Table 1.

Deep’s classification of knee deformity as it flexes from an extended position.

All Figures

thumbnail Figure 1.

The implant size and positioning of the rotational position of femoral cutting jig.

In the text
thumbnail Figure 2.

The change in coronal deformity (X axis) of knee femoro-tibial mechanical alignment angle, as the knee flexes (Y axis): first part showing before the surgery valgus deformity first increases and then decreases as the knee flexes (Type 4A in Deep’s classification) and second part showing the axis achieved after surgery of same patient (Neutral alignment).