Original Article
No dislocation and low complication rate for a modern dual mobility cup with pre-impacted femoral head in primary hip replacement: A consecutive series of 175 hips at minimum 5-year follow-up
1
Department of Orthopedic Surgery and Sport Medicine, Croix-Rousse Hospital, FIFA Medical Center of Excellence, 69004 Lyon, France
2
Université de Lyon, Université Claude Bernard Lyon 1, 69100 Lyon, France
3
The Midyorkshire Hospitals NHS Trust, WF14DG Wakefield, United Kingdom
4
Fortis Hospital, 560076 Bangalore, India
5
EA 7424, Interuniversity Laboratory of Human Movement Science, Université Lyon 1, 69100 Lyon, France
6
Université de Lyon, Université Claude Bernard Lyon 1, IFSTTAR, LBMC UMR_T9406, 69622 Lyon, France
* Corresponding author: constant.foissey@chu-lyon.fr
Received:
30
October
2022
Accepted:
26
November
2022
Introduction: Despite its excellent results in preventing dislocation, the dual mobility cup (DMC) is still struggling to be adopted by some teams due to premature wear and loosening reported on first-generation implants. Therefore, this study aimed to assess the mid-term survivorship of a modern DMC with a pre-impacted head and the radio-clinical results at a minimum follow-up of 5 years. Methods: This was a retrospective single-centre study performed on patients who had a primary total hip replacement for osteoarthritis in 2016. The cup was a third-generation DMC with a pre-impacted femoral head. Clinical (harris hip score (HHS)) and radiological (cup abduction, anteversion, overhang, and radiolucent lines) results were recorded, as well as complications, particularly dislocations and survivorship. A minimum of five years of follow-ups was required. Results: One hundred and seventy-five hips (167 patients) met the inclusion criteria. Five hips (2.9%, 5/175) were lost to follow-up and excluded from the postoperative analysis. The mean follow-up period was 70 ± 2.9 months [63.6–76.5]. Three cups needed revision surgery (1.8%, 3/170): one for septic loosening, and two for chronic infection. At 77 months, the global survival probability was 98.2% ± 1, and the survival probability excluding septic aetiology was 100%. There was a significant improvement in the HHS from pre-operatively (48.3 ± 6.0 [14.0–70.0]) to post-operatively (96 ± 4.5 [50–100]) (p < 0.0001). There were no postoperative dislocations recorded, nor any iliopsoas-impingement or symptomatic cam-effect. Discussion: This study showed excellent survival and good radiological and clinical results of this dual mobility cup at a mid-term follow-up. None of the patients had dislocation or any specific complication feared with dual mobility cups.
Key words: Total hip arthroplasty / Dual mobility cup / Survival / Dislocation / Complication / Loosening
© The Authors, published by EDP Sciences, 2023
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Introduction
The efficiency of dual mobility cups (DMC) in reducing dislocation is no longer to be proven. Since its introduction in France by Gilles Bousquet in 1974 [1], many studies have shown its efficiency in trauma, primary and revision surgeries [2–7]. These promising results were observed very soon in the original studies using first-generation DMC [8].
Nevertheless, time-specific issues revealed themselves: random long-term biological fixation due to simple alumina coating [9]; too prominent design leading to iliopsoas impingement muscle and the femoral neck [10, 11]; premature wear of the third articulation leading to intra-prosthetic dislocation (IPD) due to the absence of a chamfer, a poor quality conventional polyethylene, and a non-adapted wide and rough femoral neck [12–15]. Work on all of these factors (combined titanium and hydroxyapatite coating, ultra-high molecular weight polyethylene (UHMWPE), chamfers in the third articulation, pre-impacted femoral head and thin polished round neck) led to the introduction of second-generation DMCs in the late 1990s, with improved survival and almost eliminated IPDs [16–18]. Due to improved results, its use is increasingly widespread, especially in France, its native country (39.3% in primary surgery, 87.9% in revision surgery in 2018) [19, 20]. On the other hand, some teams and countries are still hesitant, they reserve it for particular indications and look for more robust results to extend their indications [20, 21].
In order to improve every potential default, the implants continue to evolve. That is why recently, in addition to the improvements mentioned above, a DMC with a pre-impacted head has been developed. This solution, in addition to saving operative time, handling and storage, makes it possible to secure the impaction which is done perfectly in the axis in a robotized way and thus protects the third articulation, in particular, the head capture mechanism which can be damaged if not impacted properly in the operating room [22]. This study aimed to assess the mid-term survivorship of modern dual mobility cups (DMC) with pre-impacted heads, as well as the radio-clinical results and dislocation rate at a minimum follow-up of 5 years.
Materials and methods
Patients
This was a retrospective study performed on patients operated on in 2016 with a total hip arthroplasty (THA) by a single surgeon experienced in hip surgery. The inclusion criteria were the use of a third-generation DMC with a pre-impacted femoral head in a primary total hip replacement for osteoarthritis. Revision surgeries, THA for neck fractures, cemented DMCs, and previous hip infections were excluded. The minimum follow-up required was five years.
Implants
The Quattro cementless dual-mobility cup (Lepine®, Genay, France) was used in all cases (Figure 1a). It is a third-generation cobalt-chromium alloy acetabular cup with a bilayer coating consisting of a sublayer of pure porous titanium covered by plasma-sprayed hydroxyapatite. The titanium sublayer is obtained by vacuum-plasma spraying, which prevents the interposition of damaging titanium oxides and the weakening of its cohesion with the implant or of the implant itself. The acetabular shell is a half-sphere augmented by a partial cylindrical extension, increasing the coverage angle until 12° allowing additional joint stability. The primary fixation is ensured by eight equatorial fins. The inner surface of the cup is highly polished without any holes for screws or spikes, or additional fixation devices. The polyethylene liner is machined from standard UHMWPE sterilised by ethylene oxide. It is five-eighths of a sphere and features a narrowed section in which the head is secured using a bearing press. In addition, a specific chamfer was made to decrease contact between the retention rim and the neck of the femoral component. The bearing couple was either metal-polyethylene (cobalt-chrome) for 22.2 mm head (46 and 48 mm cups) or ceramic-polyethylene (BIOLOX®delta, CeramTec GmbH, Plochingen, Germany) for 28 mm head (cups over 50 mm). All the heads were pre-impacted in the polyethylene in the factory (Figure 1b).
Figure 1 Quattro®cup, hemispherical cup with partially extended rim for additional joint stability (a). Pre-impacted head and liner, this solution, in addition to saving operative time, handling and storage, makes it possible to secure the impaction which is done perfectly in the axis in a robotized way and thus protects the third articulation, in particular, the head capture mechanism (b). |
The femoral stem used was collared, cementless with a nonporous fully hydroxyapatite coating on a forged titanium alloy stem, a proximal quadrangular cross-section and an elliptic mirror-polished neck (Targos®, Lepine, Genay, France) [23].
Surgical technique
Patients were operated on either by direct anterior approach (DAA) without any traction table [24] or by mini-invasive posterior approach [25]. In our institution, DAA is standard, except for body mass index (BMI) ≥ 40 kg/m2, complex THA requiring corrective osteotomy (e.g., congenital hip dysplasia, corrective leg length discrepancy), over 85 years old with osteoporosis, or previous femoral or pelvic osteotomy [26, 27]. DMCs are used for all patients operated by posterior approach and for patients operated by anterior approach if they were older than 65 years of age and if there was a high risk of dislocation (e.g. epilepsy, Parkinson’s disease, extensive arthrolysis, spinal fusion…).
Data collection
The demographic data (age, gender, BMI, etiologies) and a pre-operative harris hip score (HHS) were collected. Postoperative clinical assessment was evaluated using the HHS and by collecting complications by an independent observer. The results were classified as: poor (HHS < 70), fair (HHS = 70–79), good (HHS = 80–89), and excellent (HHS = 90–100) [28].
On the postoperative pelvic AP views, the measurements of implant positioning were performed by an independent observer. Cup abduction was the angle between the cup axis and the parallel between the inter-teardrop line. The anteversion was calculated using Widmer’s method [29]. The cup positioning was considered in the safe zone if abduction was between 30° and 50° and anteversion between 10° and 30° [30]. The cup overhang was evaluated on a Lequesne profile view.
Our radiological analysis also focused on socket migration and evaluation of the bone-prosthesis interface. Radiolucent lines (>2 mm) or gaps in the different areas described by DeLee and Charnley [31] were also recorded.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committees and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. For this study, formal consent was not required.
Statistical analysis
Continuous variables were described by their mean value, standard deviation and minimum and maximum values. Qualitative variables were summarized as percentages. The Kaplan-Meyer method was performed to illustrate the survival of implants. The statistical analysis was performed using XLSTAT™ software (version 2021, AddInsoft, Paris, France).
Results
Two hundred and twenty-seven THAs were performed in our institution in 2016. Sixty-two hips (26%, 62/237) were excluded as they received a standard cup. One hundred and seventy-five hips (74%, 175/237) (167 patients) met the inclusion criteria. Five hips (2.9%, 5/175) were lost to follow-up and excluded from the postoperative analysis (Figure 2). The mean follow-up period was 70 ± 2.9 months [63.6–77]. Ninety-eight hips (58%, 98/170) were operated by DAA, and 72 (42%, 72/170) by posterior approach. Demographic data and per-operative data are presented in Table 1.
Figure 2 Flowchart, THA: total hip arthroplasty, DMC: dual mobility cup. |
Pre-operative demographic data and per-operative data.
There were five infections (2.9%, 5/170): two cases (1.2%, 2/170) of early acute infection treated by DAIR (debridement, antibiotics and implant retention) and three cases (1.7%, 3/170) of chronic sepsis which resulted in one acetabular loosening that underwent a one-step bipolar revision (0.6%, 1/170) at 16.6 months and two chronic pain that underwent bipolar revision (1.2%, 2/170) at 9.7 and 25.9 months. Three other hips needed a revision surgery (1.7%, 3/170): two for aseptic femoral loosening with a unipolar revision (1.2%, 2/170) and one for peri-prosthetic fracture with femoral osteosynthesis (0.6%, 1/170). At 77 months, the global survival probability of the cup was 98.2% ± 1, survival probability of the cup excluding septic aetiology was 100% (Figure 3). There was no aseptic acetabular loosening or postoperative dislocation recorded, nor any iliopsoas-impingement or symptomatic cam-effect (Table 2).
Figure 3 Survival analysis of the acetabular cup. (a) Global and (b) excluding septic etiologies. |
Intra and postoperative complications, and postoperative functional outcomes at the last follow-up.
There was a significant improvement of the HHS from pre-operatively (48.3 ± 6.0 [14.0–70.0]) to post-operatively (96 ± 4.5 [50–100]) (p < 0.0001) (Tables 1 and 2). 90% (153/170) sustained excellent results with HHS ≥ 90 (Table 2). 90.6% (154/170) of the cups were in the safe zone in the frontal plane, 86.5% (147/170) in the transversal plane and 80% (136/170) in both planes (Figure 4). The mean inclination was 41.0 ± 5.3 [26.0–70.0], mean anteversion was 19.8 ± 6.3 [0–77.8]. Two patients had an anterior cup overhang (1.2%, 2/170) of 4 and 1 mm, with no associated iliopsoas syndrome. There was no evidence of cup migration, nor any radiolucent lines (signs of loosening), or any eccentric position of the head within the cup noted (Figure 5).
Figure 4 Cup positioning according to safe zones. 90.6% (154/170) of the cups were in the safe zone in the frontal plane, 86.5% (147/170) in the transversal plane, and 80% (136/170) in both planes. |
Figure 5 No eccentric position of the femoral head in the cup (as a sign of polyethylene wear) was found at the last follow-up. |
Discussion
The main finding of this study is that the mid-term survivorship of this modern dual mobility cup with a pre-impacted femoral head was excellent in this study (98.2% at 6.4 years). No dislocations were recorded, nor any specific complications classically described for dual mobility cups (psoas impingement, IPD, symptomatic cam effect) were seen at 70 months FU. The survival rate excluding septic complications was 100% at 6.4 years.
The first-generation DMCs, which did not have hydroxyapatite coating to promote osteointegration, were associated with an increased risk of aseptic loosening and a higher occurrence of radiolucent lines [9, 32]. Second generations of DMCs with a double coating of titanium and hydroxyapatite allow for a theoretically better osteointegration leading to a theoretically higher survival probability. Even though direct statistical comparisons are not possible, survivorship seems higher with second-generation DMC with a survival probability >97% at 5–11 years in the literature [16–18, 33–41] and in our study. The large sliding surface of DMCs is often pointed out as a source of particle release, yet an in-vitro study found with a UHMWPE that the wear performance between a DMC and a standard cup with a 28 mm ceramic head was equivalent [42]. Even though last-generation polyethylene is not already the reference in DMC, some authors recommend their use in daily practice based on finite element analysis, even obtaining with a 22.2 mm femoral head a similar amount of volumetric wear than a conventional acetabular component with a 32-mm-diameter femoral head against XLPE [22, 43]. If we consider such excellent results, we might expect that new generations of DMC enhanced by the latest surface coating (highly porous titanium) [44] and sliding surface (cross-linked polyethylene(XLPE)) [45] technologies continue to ensure optimal survival at the longest follow-up.
Optimal stability is the aim for dual mobility use in primary and revision hip replacement. There were no cases of dislocation in our series, which corroborate the literature [6, 16–18, 33–40]. These results are all the more relevant as there was a high proportion of the population which was at high risk of dislocation (36.9% of obese patients, 38.6% ≥75 years old) [46]. This protective effect on obese patients has already been described [47], as well as in elderly patients [48].
The partially extended rim of this implant allows an increased coverage angle of up to 12° participating in this anti-dislocation shield [49]. This increased coverage being only on the upper side, this increased stability was not at the expense of a rising risk of psoas impingement nor to symptomatic cam effect as those complications were not recorded in this study. Other designs can be used to face this compromise, and some manufacturers use a cylinder-spheric design (sphere augmented by a cylinder of 3 mm at the equatorial area) [38], while others use an anatomical design (anterior notch to accommodate the iliospsoas tendon, and a posterior and inferior prominence to increase the congruence) [45]. Whatever the design, the surgical technique is crucial, and appropriate cup positioning with no anterior impingement is mandatory.
Not recorded as well in this study, IPDs seem to become anecdotal. None of the recent publications following new-generation DMC found any intraprostatic dislocations (IPDs) [6]. The source of this improvement is multifactorial. It is mainly due to the improvement of the quality of the polyethylene and the improvement of the 3rd articulation (chamfer, thin polished round neck) [42]. The originality of the cup used here is that it respects the third articulation as much as possible by standardizing the impaction of the heads robotically prior to surgery in order to avoid any damage to the head capture mechanism. Even if at mid-term follow-up it is difficult to prove that the benefit of this particular construct is not only theoretical, the current study does support its safety.
Particular care has been taken to use ceramic heads as frequently as possible as they are known to give less fretting corrosion at the head–neck interface (trunionitis) [50]. By analogy with the metal on metal bearing surfaces used in the past, we also know that it is preferable to avoid metal on metal head-neck interface as much as possible when placing large-diameter heads since these are associated with an increased risk of trunionitis due to larger frictional torques and larger bending moments are generated on the femoral trunnion [51].
Even though the implementation of a monobloc DMC is known to be more difficult due to the lack of a solid anchorage point at the bottom of the cup for the holder, this study found an excellent global placement of the cup (90.1% well-positioned in the frontal plane, 86% in the transversal plane). Those results are similar to those found with standard cups [28, 30], keeping in mind that 42% of the procedures were performed through DAA with fluoroscopic control, which is known to allow for good control of the cup positioning [52].
There are some limitations to this study that need to be acknowledged. The assessment of implant position was on radiographs and not on CT scan, however, a CT scan is not recommended for routine THA follow-up. This would have exposed study patients to increased radiation and would not reflect common practice. Another limitation is the mid-term follow-up, knowing that the ideal follow-up to study loosening would be after 10–15 years, but this is the first report of the pre-impacted femoral head in modern DMC and considering the mean age of 72 years in our series, the mean FU of 6 years is relevant.
This study has several advantages, including a large cohort size with almost no loss at follow-up of a minimal 5y with the same implant and the same surgeon. Our results at mid-term FU seem to support the routine use of DMC in primary hip replacement for the patient at risk of instability since the stability is achieved without specific complications related to the implant.
Conclusion
This study showed an excellent survival of this modern dual mobility cup at a mid-term follow-up. No dislocations were recorded, nor any specific complications (intraprosthetic dislocation, psoas impingement, symptomatic cam effect), which are feared with dual mobility cups were seen. The authors would recommend using dual mobility implants for primary hip replacement in patients at risk of instability.
Conflicts of interest
No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. CF, CB, AP and VR declare that they have no confict of interest. ES: Consultant for Corin. SL: Consultant for Stryker, Smith Nephew, Heraeus, Depuy Synthes; Institutional research support from Groupe Lepine, Amplitude; Editorial Board for Journal of Bone and Joint Surgery (Am).
Ethics approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The Advisory Committee on Research Information Processing in the Field of Health (CCTIRS) approved this study on June 4, 2015 under number 15-430. For this type of study formal consent is not required.
Funding
There is no funding source.
Design
A retrospective observational study, level IV.
Availability of data and material
Not applicable.
Code availability
Not applicable.
Authors’ contribution
Constant Foissey: Study design, data collection, statistical analysis, literature review and manuscript writing. Cécile Batailler: Study design, manuscript editing. Aditya B.J. Premkumar: Data collection. Vishal Rajput: manuscript editing. Elvire Servien: Study design, manuscript editing. Sébastien Lustig: Study design, supervision, literature review and manuscript editing. All authors read and approved the final manuscript.
Acknowledgments
Nothing to declare.
References
- Noyer D, Caton JH (2017) Once upon a time. Dual mobility: History. Int Orthop 41, 611–618. [CrossRef] [PubMed] [Google Scholar]
- Alberio RL, Rusconi M, Martinetti L, et al. (2021) Total hip arthroplasty (THA) for femoral neck fractures: Comparison between standard and dual mobility implants. Geriatrics (Basel) 6, 70. [CrossRef] [PubMed] [Google Scholar]
- Batailler C, Fary C, Batailler P, et al. (2017) Total hip arthroplasty using direct anterior approach and dual mobility cup: Safe and efficient strategy against post-operative dislocation. Int Orthop 41, 499–506. [CrossRef] [PubMed] [Google Scholar]
- Hoskins W, McDonald L, Claireaux H, et al. (2022) Dual-mobility constructs versus large femoral head bearings in primary and revision total hip arthroplasty: A systematic review and meta-analysis of comparative studies. Hip Int 11207000221082928. https://doi.org/10.1177/11207000221082927. [Google Scholar]
- Schmidt A, Batailler C, Fary C, et al. (2020) Dual mobility cups in revision total hip arthroplasty: Efficient strategy to decrease dislocation risk. J Arthroplasty 35, 500–507. [CrossRef] [PubMed] [Google Scholar]
- Batailler C, Fary C, Verdier R, et al. (2017) The evolution of outcomes and indications for the dual-mobility cup: A systematic review. Int Orthop 41, 645–659. [CrossRef] [PubMed] [Google Scholar]
- Romagnoli M, Grassi A, Costa GG, et al. (2019) The efficacy of dual-mobility cup in preventing dislocation after total hip arthroplasty: A systematic review and meta-analysis of comparative studies. Int Orthop 43, 1071–1082. [CrossRef] [PubMed] [Google Scholar]
- Leclercq S, El Blidi S, Aubriot JH (1995) Traitement de la luxation récidivante de prothèse totale de hanche par le cotyle de Bousquet A propos de 13 cas. Rev Chir Orthop Reparatrice Appar Mot 81, 389–394. [PubMed] [Google Scholar]
- Philippot R, Farizon F, Camilleri J-P, et al. (2008) Survival of dual mobility socket with a mean 17 years follow-up. Rev Chir Orthop Reparatrice Appar Mot 94, 43–48. [CrossRef] [PubMed] [Google Scholar]
- Epinette J-A, Harwin SF, Rowan FE, et al. (2017) Early experience with dual mobility acetabular systems featuring highly cross-linked polyethylene liners for primary hip arthroplasty in patients under fifty five years of age: An international multi-centre preliminary study. Int Orthop 41, 543–550. [CrossRef] [PubMed] [Google Scholar]
- Chughtai M, Mistry JB, Diedrich AM, et al. (2016) Low frequency of early complications with dual-mobility acetabular cups in cementless primary THA. Clin Orthop Relat Res 474, 2181–2187. [CrossRef] [PubMed] [Google Scholar]
- Farizon F, Baulot E, Fessy MH (2018) Historique: Des erreurs du passé à une double mobilité moderne. In: La double mobilité en marche dans les prothèses totales de hanche, Elsevier Masson. pp. 3–6. [Google Scholar]
- Lecuire F, Benareau I, Rubini J, Basso M (2004) Intra-prosthetic dislocation of the Bousquet dual mobility socket. Rev Chir Orthop Reparatrice Appar Mot 90, 249–255. [CrossRef] [PubMed] [Google Scholar]
- Neri T, Boyer B, Batailler C, et al. (2020) Dual mobility cups for total hip arthroplasty: Tips and tricks. Sicot J 6, 17. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
- Philippot R, Boyer B, Farizon F (2013) Intraprosthetic dislocation: A specific complication of the dual-mobility system. Clin Orthop Relat Res 471, 965–970. [CrossRef] [PubMed] [Google Scholar]
- Vermersch T, Viste A, Desmarchelier R, Fessy M-H (2015) Prospective longitudinal study of one hundred patients with total hip arthroplasty using a second-generation cementless dual-mobility cup. Int Orthop 39, 2097–2101. [CrossRef] [PubMed] [Google Scholar]
- Prudhon J-L, Ferreira A, Verdier R (2013) Dual mobility cup: Dislocation rate and survivorship at ten years of follow-up. Int Orthop 37, 2345–2350. [CrossRef] [PubMed] [Google Scholar]
- Gaillard R, Kenney R, Delalande J-L, et al. (2019) Ten- to 16-year results of a modern cementless dual-mobility acetabular implant in primary total hip arthroplasty. J Arthroplasty 34, 2704–2710. [CrossRef] [PubMed] [Google Scholar]
- Caton JH, Ferreira A (2017) Dual-mobility cup: A new French revolution. Int Orthop 41, 433–437. [CrossRef] [PubMed] [Google Scholar]
- Castiello E, Moghnie A, Tigani D, Affatato S (2022) Dual mobility cup in hip arthroplasty: An in-depth analysis of joint registries. Artif Organs 46, 804–812. [CrossRef] [PubMed] [Google Scholar]
- Horriat S, Haddad FS (2018) Dual mobility in hip arthroplasty: What evidence do we need? Bone Joint Res 7, 508–510. [CrossRef] [PubMed] [Google Scholar]
- Malatray M, Roux J-P, Gunst S, et al. (2017) Highly crosslinked polyethylene: A safe alternative to conventional polyethylene for dual mobility cup mobile component. A biomechanical validation. Int Orthop 41, 507–512. [CrossRef] [PubMed] [Google Scholar]
- Batailler C, Fary C, Servien E, Lustig S (2018) Influence of femoral broach shape on stem alignment using anterior approach for total hip arthroplasty: A radiologic comparative study of 3 different stems. PLoS One 13, e0204591. [CrossRef] [PubMed] [Google Scholar]
- Lustig S (2015) Anterior approach (without specific table) and dual mobility acetabular component. Maitrise Orthop 243, 1–5. [Google Scholar]
- Aït Si Selmi T, Lustig S, Dojcinovic S, Neyret P (2006) Morbidity and reliability of total hip implants positioning using the posterior minimally invasive approach: A consecutive series of 100 cases. Rev Chir Orthop Reparatrice Appar Mot 92, 752–759. [CrossRef] [PubMed] [Google Scholar]
- Foissey C, Fauvernier M, Fary C, et al. (2020) Total hip arthroplasty performed by direct anterior approach – Does experience influence the learning curve? Sicot J 6, 15. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
- Foissey C, Kenney R, Luceri F, et al. (2021) Greater trochanter fractures in the direct anterior approach: Evolution during learning curve, risk factors and consequences. Arch Orthop Trauma Surg 141, 675–681. [CrossRef] [PubMed] [Google Scholar]
- Bauwens PH, Fary C, Servien E, et al. (2020) Early low complication rate of ceramic-on-ceramic total hip arthroplasty by direct anterior approach. Sicot J 6, 6p. [CrossRef] [EDP Sciences] [PubMed] [Google Scholar]
- Widmer K-H (2004) A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplasty 19, 387–390. [CrossRef] [PubMed] [Google Scholar]
- Foissey C, Fary C, Luceri F, et al. (2020) Transitioning the total hip arthroplasty (THA) technique from posterior approach in lateral position to direct anterior approach in supine position – risk factors for acetabular malpositioning and the learning curve. Int Orthop 44, 1669–1676. [CrossRef] [PubMed] [Google Scholar]
- DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121, 20–32. [Google Scholar]
- Farizon F, de Lavison R, Azoulai JJ, Bousquet G (1998) Results with a cementless alumina-coated cup with dual mobility. Int Orthop 22, 219–224. [CrossRef] [PubMed] [Google Scholar]
- Bauchu P, Bonnard O, Cyprès A, et al. (2008) The dual-mobility POLARCUP: First results from a multicenter study. Orthopedics 31. https://orthosupersite.com/view.asp?rID=37180. [Google Scholar]
- Leclercq S, Benoit J-Y, de Rosa J-P, et al. (2008) Results of the Evora dual-mobility socket after a minimum follow-up of five years. Rev Chir Orthop Reparatrice Appar Mot 94, e17–e22. [CrossRef] [PubMed] [Google Scholar]
- Vielpeau C, Lebel B, Ardouin L, et al. (2011) The dual mobility socket concept: Experience with 668 cases. Int Orthop 35, 225–230. [CrossRef] [PubMed] [Google Scholar]
- Hamadouche M, Arnould H, Bouxin B (2012) Is a cementless dual mobility socket in primary THA a reasonable option? Clin Orthop Relat Res 470, 3048–3053. [CrossRef] [PubMed] [Google Scholar]
- Combes A, Migaud H, Girard J, et al. (2013) Low rate of dislocation of dual-mobility cups in primary total hip arthroplasty. Clin Orthop Relat Res 471, 3891–3900. [CrossRef] [PubMed] [Google Scholar]
- Fessy M-H, Jacquot L, Rollier J-C, et al. (2019) Midterm clinical and radiographic outcomes of a contemporary monoblock dual-mobility cup in uncemented total hip arthroplasty. J Arthroplasty 34, 2983–2991. [CrossRef] [PubMed] [Google Scholar]
- Chouteau J, Rollier J-C, Bonnin MP, et al. (2020) Absence of instabilities and intra-prosthetic dislocations at 7 to 11 years following THA using a fourth-generation cementless dual mobility acetabular cup. J Exp Orthop 7, 51. [CrossRef] [PubMed] [Google Scholar]
- Darrith B, Courtney PM, Della Valle CJ (2018) Outcomes of dual mobility components in total hip arthroplasty: A systematic review of the literature. Bone Joint J 100-B, 11–19. [CrossRef] [PubMed] [Google Scholar]
- Caton JH, Prudhon JL, Ferreira A, et al. (2014) A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess whether a dual mobility cup has a decreased dislocation risk. Int Orthop 38, 1125–1129. [CrossRef] [PubMed] [Google Scholar]
- Gaudin G, Ferreira A, Gaillard R, et al. (2017) Equivalent wear performance of dual mobility bearing compared with standard bearing in total hip arthroplasty: In vitro study. Int Orthop 41, 521–527. [CrossRef] [PubMed] [Google Scholar]
- Wegrzyn J, Antoniadis A, Sarshari E, et al. (2022) Polyethylene wear of dual mobility cups: A comparative analysis based on patient-specific finite element modeling. Int Orthop 46, 779–787. [CrossRef] [PubMed] [Google Scholar]
- Epinette J-A, Coulomb R, Pradel S, Kouyoumdjian P (2022) Do modular dual mobility cups offer a reliable benefit? Minimum 5-year follow-up of 102 cups. J Arthroplasty 37, 910–916. [CrossRef] [PubMed] [Google Scholar]
- Epinette J-A, Béracassat R, Tracol P, et al. (2014) Are modern dual mobility cups a valuable option in reducing instability after primary hip arthroplasty, even in younger patients? J Arthroplasty 29, 1323–1328. [CrossRef] [PubMed] [Google Scholar]
- Kim Y, Morshed S, Joseph T, et al. (2006) Clinical impact of obesity on stability following revision total hip arthroplasty. Clin Orthop Relat Res 453, 142–146. [CrossRef] [PubMed] [Google Scholar]
- Maisongrosse P, Lepage B, Cavaignac E, et al. (2015) Obesity is no longer a risk factor for dislocation after total hip arthroplasty with a double-mobility cup. Int Orthop 39, 1251–1258. [CrossRef] [PubMed] [Google Scholar]
- Belgaïd V, Viste A, Fessy M-H (2022) Cementless hydroxyapatite-coated stem with dual mobility and posterior approach in over-80 year-old patients with osteoarthritis: Rates of dislocation and periprosthetic fracture at a mean 8 years’ follow-up. Orthop Traumatol Surg Res 108, 103196. [CrossRef] [PubMed] [Google Scholar]
- Affatato S, Castiello E, Amendola L, et al. (2020) Revision of a monoblock metal-on-metal cup using a dual mobility component: Is it a reasonable option? Materials (Basel) 13, E2040. [CrossRef] [PubMed] [Google Scholar]
- Hallab NJ, Messina C, Skipor A, Jacobs JJ (2004) Differences in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements. J Orthop Res, 22, 250–259. [CrossRef] [PubMed] [Google Scholar]
- Cooper HJ, Della Valle CJ (2014) Large diameter femoral heads. Bone Joint J 96-B, 23–26. [CrossRef] [PubMed] [Google Scholar]
- Kamara E, Robinson J, Bas MA, et al. (2017) Adoption of robotic vs fluoroscopic guidance in total hip arthroplasty: Is acetabular positioning improved in the learning curve? J Arthroplasty 32, 125–130. [CrossRef] [PubMed] [Google Scholar]
Cite this article as: Foissey C, Batailler C, Rajput V, Premkumar ABJ, Servien E & Lustig S (2023) No dislocation and low complication rate for a modern dual mobility cup with pre-impacted femoral head in primary hip replacement: A consecutive series of 175 hips at minimum 5-year follow-up. SICOT-J 9, 1
All Tables
Intra and postoperative complications, and postoperative functional outcomes at the last follow-up.
All Figures
Figure 1 Quattro®cup, hemispherical cup with partially extended rim for additional joint stability (a). Pre-impacted head and liner, this solution, in addition to saving operative time, handling and storage, makes it possible to secure the impaction which is done perfectly in the axis in a robotized way and thus protects the third articulation, in particular, the head capture mechanism (b). |
|
In the text |
Figure 2 Flowchart, THA: total hip arthroplasty, DMC: dual mobility cup. |
|
In the text |
Figure 3 Survival analysis of the acetabular cup. (a) Global and (b) excluding septic etiologies. |
|
In the text |
Figure 4 Cup positioning according to safe zones. 90.6% (154/170) of the cups were in the safe zone in the frontal plane, 86.5% (147/170) in the transversal plane, and 80% (136/170) in both planes. |
|
In the text |
Figure 5 No eccentric position of the femoral head in the cup (as a sign of polyethylene wear) was found at the last follow-up. |
|
In the text |
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.