Introduction

Total Knee Arthroplasty (TKA) is a common surgery performed on 1 million patients per year worldwide [1]. The dissatisfaction rate remains high, up to 30% of patients, depending on the series [2]. Outcomes depend on factors such as alignment strategy, implant positioning accuracy, implant design, and postoperative rehabilitation [2].

Knee arthroplasty aims to restore alignment of the lower limb at the Hip-Knee-Ankle (HKA) angle. Deviations greater than 3° from the intended postoperative value are considered outliers and may negatively affect mechanical and functional results [3]. Postoperative HKA depends on the precision of bone cuts and implant positioning.

Inverse kinematic alignment (iKA) is a TKA implantation technique designed to restore the pre-arthritic proximal tibial functional mechanical angle as closely as possible. Described in 2020 by P. Winnock de Grave et al., this involves a tibial cut respecting joint-line obliquity, followed by femoral adjustments to balance flexion and extension gaps, ensuring comparable residual laxity in both medial and lateral compartments [4]. This method requires providing a checked alignment, particularly in varus, in order to reduce the incidence of aseptic loosening, as suggested by Tueckin et al. [5].

Robot-assisted total knee arthroplasty (RA-TKA) surgery is a fast-growing field, offering modeling, navigation, and mechanical assistance, helping to improve the precision of bone cuts, implant positioning, and consequently the achievement of preoperatively defined alignment goals [6]. RA-TKA allows preservation of soft tissue and decreases intraoperative blood loss [7]. The Stryker MAKO™ Robot System (MAKO SmartRobotics™, Stryker Corporation, Kalamazoo, Michigan, USA), based on CT data, has been available since 2016 and has proven its effectiveness [8].

RA-TKA appears to improve radiological and functional results, compared with conventional TKA (C-TKA). Studies show a significant improvement in immediate post-operative radiological alignment with RA-TKA [9, 10]. A trend towards better functional scores also appears to be emerging [11, 12]. However, these studies contain numerous biases and do not allow us to formally settle on superiority.

The primary aim of our study is to evaluate the accuracy of the RA-TKA in achieving the iKA alignment goals defined pre-operatively through radiological assessment at 12 months post-operatively. Our secondary objectives are to evaluate clinical and functional outcomes and complication rates at 12 months post-operatively.

Materials and methods

Study design and study population

We conducted a single-center retrospective cohort study including all consecutive primary RA-TKA procedures performed between January 1, 2020, and December 31, 2022, using the MAKO Stryker™ robotic system combined with the inverse kinematic alignment technique at our university hospital. This comprises patients presenting with marked varus or valgus deformities, allowing for reflection on real-world clinical practice. Patients under 18 years or under legal protection were excluded.

Surgical procedure

All interventions were performed using the MAKO Stryker™ robot-assisted surgery system by four experienced operators. We combined this with the inverse kinematic alignment (iKA) technique, similar to functional alignments, as described by P. Winnock de Grave, with the addition of a safety zone for the proximal mechanical angle of the tibial implant at 3° varus to 2° valgus [4, 5, 1214]. The posterior-stabilized Stryker™ Triathlon implant was used.

The positioning of the femoral component was adjusted primarily by modifying its rotation to achieve identical spaces and residual ligament laxity at the internal and external compartments during both flexion and extension. This space balance was achieved using the navigation system of the MAKO Stryker™ system. A safety zone for the HKA angle was provided, ranging from 174° to 183°. The HKA angle was calculated preoperatively using the MAKO Stryker™ system, then adjusted intraoperatively.

Data collection

At the end of the procedure, the HKA angle was measured using the MAKO Stryker™ system. This value (HKA-D0) corresponded to the final alignment obtained intraoperatively in the supine, non–weight-bearing position and was automatically calculated by the navigation system after bone cuts and implant positioning.

All patients were seen at 12-month follow-up. The HKA angle at 12 months (HKA-M12) was measured on a weight-bearing full-length standing hip–knee–ankle radiograph. Coronal alignment was measured using dedicated digital planning software on calibrated images, according to the mechanical axis definition (center of the femoral head to the center of the ankle joint). Range of motion, as well as the KSS-2011 and KOOS scores assessing physical abilities and residual functional limitations, were collected at the time of postoperative consultation [15, 16]. Patients who had not completed the functional outcome questionnaire were contacted by phone to obtain the missing data. Complications assessed included surgical site infection, capsular rupture, stiffness requiring mobilization under anesthesia, or arthrolysis.

Our primary outcome, the accuracy of implant positioning, was evaluated by the rate of outliers at 12 months postoperatively. Outliers were defined as a deviation of the HKA angle at 12 months strictly greater than 3° compared to the HKA angle measured intraoperatively, i.e., the absolute difference between the HKA-D0 and HKA-M12 angles as follows:

DeltaHKA=(HKAM12)(HKAD0)Mathematical equation: $$ \mathrm{DeltaHKA}=\left(\mathrm{HKA}-\mathrm{M}12\right)\hbox{---} \left(\mathrm{HKA}-\mathrm{D}0\right) $$

The secondary outcomes included:

  • Mean difference between alignment planning and HKA measurement at 12 months.

  • KOOS and KSS-2011 scores at 12 months.

  • Knee range of motion in flexion at 12 months.

Safety outcomes analyzed were the incidence rate of the following postoperative complications:

  • Surgical site infection.

  • Non-septic incidents requiring surgical revision.

  • Mobilization or arthrolysis under general anesthesia within 12 months postoperatively: maneuvers under anesthesia are indicated in cases of joint stiffness with a flexion of less than 90° on the 45th postoperative day.

Statistical analysis

Statistical analyses were performed using Python 3 with Pandas, SciPy, NumPy [17]. Descriptive statistics were reported for all variables. Quantitative variables are expressed as means ± standard deviations, and qualitative data are presented as counts and percentages (%).

Results

From January 1, 2020, to December 31, 2022, a total of 500 arthroplasties were performed using the technique under investigation and were included in the cohort. Thirteen patients did not complete the 12-month postoperative radiographic follow-up and were excluded from the analysis (Figure 1). Patient characteristics are presented in Table 1.

Thumbnail: Figure 1 Refer to the following caption and surrounding text. Figure 1

Flowchart.

Table 1

Preoperative characteristics of the study population.

Analysis of the collected data revealed an outlier rate of 2.05% (n = 10/487) at 12 months postoperatively.

The mean HKA deviation at 12 months was 1.3° (SD ± 1.2°). The distribution of the DeltaHKA angle deviation is illustrated in Figure 2. The functional scores obtained were 214/255 (SD ± 31) for the KSS-2011 and 82/100 (SD ± 14) for the KOOS. Functional outcome data were available for 335 and 333 patients for the KSS-2011 and KOOS, respectively, as some patients declined to complete the questionnaires during follow-up visits or telephone interviews. At 12 months postoperatively, the mean flexion range of motion was 127° (SD ± 11°). Detailed results are presented in Table 2.

Thumbnail: Figure 2 Refer to the following caption and surrounding text. Figure 2

Results: Distribution of DeltaHKA.****DeltaHKA as defined: (HKA-M12) – (HKA-D0). ****1 point = 1 event. Outliers shown in red color.

Table 2

Results.

Among the 10 outlier cases, 6 patients presented with major preoperative varus deformities greater than 10°, and 4 patients had obesity (BMI > 30 kg/m2). No corrective intervention related to coronal alignment was required during the first postoperative year. The only secondary procedure performed was one manipulation under general anesthesia for postoperative stiffness. Notably, 7 of the 10 outliers exhibited a valgus alignment at the 1-year follow-up, despite all having presented with a preoperative varus deformity. Subgroup analysis comparing outliers and non-outliers demonstrated no statistically significant differences in KSS-2011 scores (p = 0.108), KOOS scores (p = 0.290), or maximal flexion at one year (p = 0.189).

Twenty-three complications occurred, including 6 cases of sepsis and 11 cases of joint stiffness requiring surgical intervention. All adverse event data are presented in Table 3.

Table 3

Safety outcomes.

Discussion

The primary aim was to assess RA-TKA accuracy in achieving preoperative alignment targets by analyzing 12-month HKA outliers. RA-TKA showed excellent implantation accuracy with a low outlier rate (2.05%) and minimal HKA deviation (mean DeltaHKA 1.3°).

The cohort demonstrates a very low rate of HKA outliers. The robot-assisted system reduces inaccuracies associated with conventional mechanical alignment techniques. The real-time navigation system allows continuous intraoperative monitoring of alignment and verification of final implant positioning, guaranteeing its fidelity to the intended plan. Ligament balancing is optimized, preventing overpressure in one compartment that could lead to prosthetic subsidence. Even within the ±3° range, postoperative HKA deviation averaged 1.3°, reflecting stable outcomes.

In the literature, outlier rates for C-TKA range from 13% to 38% [9, 10]. Our findings support RA-TKA use for improved postoperative alignment. Previous studies on this topic were affected by biases such as heterogeneous surgical protocols and limited statistical power. They reported a very low proportion of outliers for RA-TKA.

HKA outliers are associated with poorer mechanical and functional outcomes [3, 13]. Comparing functional and clinical scores between RA-TKA and C-TKA remains a key topic. We assessed recovery by assessing the KSS-2011 and KOOS scores, with total scores of 214/255 and 82/100, respectively, reflecting highly satisfactory functional outcomes at 12 months. Given the primarily radiographic focus of our study, preoperative functional outcome data were not collected, which prevented us from evaluating changes in functional outcomes between the preoperative and postoperative periods. Prior studies had heterogeneous assessments and rarely used the most recent KSS-2011 score [15].

A previous small-scale study investigated RA-TKA combined with iKA [12]. It reported encouraging functional scores but lacked statistical power. A meta-analysis by Zhang et al. indicates RA-TKA superiority in postoperative KSS scores, though heterogeneity in surgical protocols and robotic systems limits direct applicability [18]. Overall, RA-TKA appears to provide better functional outcomes, but evidence on the combined RA-TKA + iKA technique remains limited.

We observed an overcorrection toward valgus alignment in most outlier cases; this pattern may be related to the management of severe varus knees. In these cases, minimal medial tibial resection is performed. Implant fixation may therefore rely on a predominantly sclerotic bone surface within the medial compartment. Subsequent medial tibial settling could progressively modify coronal limb alignment and contribute to the development of a more valgus alignment than that achieved intraoperatively. However, this hypothesis remains speculative and could not be formally assessed in the present study. Although no significant difference in functional outcomes was observed at short-term follow-up between aligned cases and outliers, this finding does not preclude potential long-term implications. Deviation from the intended intraoperative coronal alignment target may alter load distribution across the knee joint and could influence implant fixation over time. Longer follow-up is therefore necessary to determine whether such alignment differences impact implant durability [19].

Postoperative complications were low: surgical site infection rate 1.2% at 1 year, comparable to 1.5–2% for C-TKA, suggesting non-inferiority [2022]. Joint stiffness requiring intervention occurred in 2.3% of cases; prior reports for C-TKA range from 2.5 to 4.8%, and less than 2% for RA-TKA [23, 24]. The improvement observed in the RA-TKA group may be explained by more precise intraoperative assessment of ligamentous laxity as well as guided bone preparation, allowing the achievement of patient-specific implant positioning targets. The combination of iKA and MAKO™ robotic navigation facilitates satisfactory ligament balancing, proper joint balance, and peripheral isometric deformation both internally and externally.

This study has several limitations. Although the 12-month follow-up provides relevant short-term radiographic and functional data, it remains shorter than the 2-year follow-up commonly reported in arthroplasty studies and is limited compared with the expected lifespan of a knee implant. While sufficient to assess early outcomes, this follow-up duration may not capture longer-term phenomena such as alignment drift, implant loosening, late complications, or the evolution and stabilization of functional outcomes beyond the first postoperative year. Early mechanical and functional improvements may persist over time, but this remains to be demonstrated. Longer-term follow-up investigations are currently underway at our institution and are expected to provide further insight into the durability of the radiographic and clinical outcomes.

The comparison between intraoperative and 12-month HKA values must be interpreted cautiously. Intraoperative alignment was recorded in the supine, non–weight-bearing position using the navigation system, whereas the 12-month measurements were obtained from standing weight-bearing radiographs. A systematic discrepancy between intraoperative MAKO™ measurements and postoperative weight-bearing radiographs may partly explain our findings. Glowalla et al. reported that postoperative full-leg standing radiographs consistently demonstrated less residual varus alignment than intraoperative MAKO™ measurements, with a mean difference of 1.8° between the two assessment methods, which could bias our findings [25]. These findings suggest that the robotic navigation system may tend to overestimate residual varus deformity intraoperatively. However, unlike Glowalla et al., our study did not include paired intraoperative and postoperative alignment measurements, preventing us from directly assessing the presence and magnitude of such a systematic measurement bias in our cohort.

As this study was observational, postoperative imaging followed the institutional standard protocol. This protocol does not include systematic early postoperative weight-bearing radiographs (day 1–2) or postoperative CT scans. Such additional imaging could have enabled comparison with the intraoperative measurement. However, it was not performed as it would have implied additional radiation exposure and increased cost beyond routine clinical care.

The study cohort is single-center, which may limit external validity. Finally, this study is observational and non-comparative; it does not allow definitive conclusions regarding the superiority of RA-TKA combined with iKA over C-TKA.

On the other hand, our cohort presents significant advantages. The study population is large, encompassing several hundred arthroplasties performed using a homogeneous protocol, a number rarely reached in the current literature [9, 10]. The surgical team performing the procedures consisted of experts, and the MAKO Stryker™ robot used has demonstrated efficacy and reliability. RA-TKA is rapidly expanding, and inverse kinematic alignment shows encouraging results. The combination of these two techniques appears to be a promising approach for TKA. To our knowledge, no study of this scale has investigated their combined use.

To conclude, RA-TKA combined with iKA enables the achievement of preoperatively defined alignment targets with excellent precision and a low 12-month outlier rate. These high-quality observational data highlight postoperative functional scores and joint ranges of motion, reflecting good recovery of the operated limb at 12 months. The alignment findings appear favorable in the context of published data, although the present study does not permit direct comparison.

Funding

This research did not receive any specific funding.

Conflicts of interest

The authors declare that they have no relevant financial or non-financial interests to report.

Data availability statement

Data associated with this article cannot be disclosed due to legal reasons.

Author contribution statement

B. Bertot: Conceptualization, Methodology, Investigation, Writing original draft.

A. Royet: Conceptualization, review and editing.

A. Stordeur: Statistical analysis.

P-H. Vermorel: Review and editing.

F. Farizon: Review and editing.

R. Philippot: Supervision, study design, manuscript review.

Ethics approval

Ethical approval was not required.

References

  1. Shan L, Shan B, Suzuki A, et al. (2015) Intermediate and long-term quality of life after total knee replacement: a systematic review and meta-analysis. J Bone Joint Surg Am 97, 156–168. [Google Scholar]
  2. Canovas F, Dagneaux L (2018) Quality of life after total knee arthroplasty. Orthop Traumatol Surg Res 104, S41–S46. [CrossRef] [PubMed] [Google Scholar]
  3. Hadi M, Barlow T, Ahmed I, et al. (2015) Does malalignment affect revision rate in total knee replacements: a systematic review of the literature. Springerplus 4, 835. [Google Scholar]
  4. Winnock de Grave P, Kellens J, Luyckx T, et al. (2022) Inverse kinematic alignment for total knee arthroplasty. Orthop Traumatol Surg Res 108, 103305. [Google Scholar]
  5. Tuecking L-R, Savov P, Windhagen H, Ettinger M (2020) Current evidence base for kinematic alignment. Orthopade 49, 584–592. [Google Scholar]
  6. Mason JB, Fehring TK, Estok R, et al. (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 22, 1097–1106. [CrossRef] [PubMed] [Google Scholar]
  7. Kayani B, Konan S, Pietrzak JRT, Haddad FS (2018) Iatrogenic bone and soft tissue trauma in robotic-arm assisted total knee arthroplasty compared with conventional jig-based total knee arthroplasty: A prospective cohort study and validation of a new classification system. J Arthroplasty 33, 2496–2501. [Google Scholar]
  8. Batailler C, Fernandez A, Swan J, et al. (2021) MAKO CT-based robotic arm-assisted system is a reliable procedure for total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc 29, 3585–3598. [CrossRef] [PubMed] [Google Scholar]
  9. Song E-K, Seon J-K, Yim J-H, et al. (2013) Robotic-assisted TKA reduces postoperative alignment outliers and improves gap balance compared to conventional TKA. Clin Orthop Rel Res 471, 118. [Google Scholar]
  10. Liow MHL, Xia Z, Wong MK, et al. (2014) Robot-assisted total knee arthroplasty accurately restores the joint line and mechanical axis. A prospective randomised study. J Arthroplasty 29, 2373–2377. [Google Scholar]
  11. Liow MHL, Goh GS-H, Wong MK, et al. (2017) Robotic-assisted total knee arthroplasty may lead to improvement in quality-of-life measures: a 2-year follow-up of a prospective randomized trial. Knee Surg Sports Traumatol Arthrosc 25, 2942–2951. [CrossRef] [PubMed] [Google Scholar]
  12. Winnock de Grave P, Kellens J, Tampere T, et al. (2023) Clinical outcomes in TKA are enhanced by both robotic assistance and patient specific alignment: a comparative trial in 120 patients. Arch Orthop Trauma Surg 143, 3391–3399. [Google Scholar]
  13. Ritter MA, Davis KE, Meding JB, et al. (2011) The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am 93, 1588–1596. [CrossRef] [PubMed] [Google Scholar]
  14. Kim Y-H, Park J-W, Kim J-S, Park S-D (2014) The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthop 38, 379–385. [Google Scholar]
  15. Scuderi GR, Bourne RB, Noble PC, et al. (2012) The new knee society knee scoring system. Clin Orthop Rel Res® 470, 3. [Google Scholar]
  16. Roos EM, Lohmander LS (2003) The Knee injury and Osteoarthritis Outcome Score (KOOS): From joint injury to osteoarthritis. Health Qual Life Outcome 1, 64. [Google Scholar]
  17. Van Rossum G, Drake FL (2009) Python 3 Reference Manual. CreateSpace, Scotts Valley, CA. [Google Scholar]
  18. Zhang J, Ndou WS, Ng N, et al. (2022) Robotic-arm assisted total knee arthroplasty is associated with improved accuracy and patient reported outcomes: a systematic review and meta-analysis. Knee Surg Sports Traumatol Arthrosc 30, 2677–2695. [CrossRef] [PubMed] [Google Scholar]
  19. Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res, 153–156. [Google Scholar]
  20. Roth VR, Mitchell R, Vachon J, et al. (2017) Periprosthetic infection following primary hip and knee arthroplasty: The impact of limiting the postoperative surveillance period. Infect Control Hosp Epidemiol 38, 147–153. [Google Scholar]
  21. Weinstein EJ, Stephens-Shields AJ, Newcomb CW, et al. (2023) Incidence, microbiological studies, and factors associated with prosthetic joint infection after total knee arthroplasty. JAMA Netw Open 6, e2340457. [Google Scholar]
  22. Cahill JL, Shadbolt B, Scarvell JM, Smith PN (2008) Quality of life after infection in total joint replacement. J Orthop Surg (Hong Kong) 16, 58–65. [Google Scholar]
  23. Zhang J, Matzko CN, Sawires A, et al. (2022) Adoption of robotic-arm-assisted total knee arthroplasty is associated with decreased use of articular constraint and manipulation under anesthesia compared to a manual approach. J Knee Surg 35, 849–857. [Google Scholar]
  24. Ofa SA, Ross BJ, Flick TR, et al. (2020) Robotic total knee arthroplasty vs conventional total knee arthroplasty: A Nationwide database study. Arthroplast Today 6, 1001–1008.e3. [Google Scholar]
  25. Glowalla C, Langer S, Lenze U, et al. (2023) Postoperative full leg radiographs exhibit less residual coronal varus deformity compared to intraoperative measurements in robotic arm-assisted total knee arthroplasty with the MAKOTM system. Knee Surg Sports Traumatol Arthrosc 31, 3912–3918. [CrossRef] [PubMed] [Google Scholar]

Cite this article as: Bertot B, Royet A, Stordeur A, Farizon F, Vermorel PH & Philippot R (2026) Accuracy and functional outcomes of robot-assisted total knee arthroplasty combined with inverse kinematic alignment: a 500-case cohort study. SICOT-J 12, 44. https://doi.org/10.1051/sicotj/2026043.

All Tables

Table 1

Preoperative characteristics of the study population.

Table 3

Safety outcomes.

All Figures

Thumbnail: Figure 1 Refer to the following caption and surrounding text. Figure 1

Flowchart.

In the text
Thumbnail: Figure 2 Refer to the following caption and surrounding text. Figure 2

Results: Distribution of DeltaHKA.****DeltaHKA as defined: (HKA-M12) – (HKA-D0). ****1 point = 1 event. Outliers shown in red color.